Thank you for purchasing a Sky
Quality Meter (SQM) from
Unihedron!
Features
The SQM has the following features:
It is sensitive only to visual light
(there is a near-infrared blocking
filter in front of the sensor).
The effects of temperature on the
“dark frequency" of the sensor are
removed.
The effects of temperature on the
microcontroller oscillator are
removed.
It is protected against accidental
reversal of battery polarity.
Each SQM is calibrated using a NIST-
traceable light meter. The absolute
precision of each meter is believed to
be ±10% (±0.10 mag/arcsec²). The
difference in zeropoint between each
calibrated SQM is typically ±10%
(±0.10 mag/sq arcsec)
The brightness of the numeric LED
display has two (automatic) settings.
Under dark skies, you won't have your
dark adaption ruined by use of your
SQM! Under urban skies, the display
will be correspondingly brighter.
A repeating audible beep indicates
when a measurement is in progress.
Any kind of 9V battery is usable. The
SQM contains a voltage regulator to
power the sensor, microcontroller and
other components.
After reading is taken and displayed,
the meter automatically turns itself off.
The Half Width Half Maximum
(HWHM) of the angular sensitivity is
~42°.
The unit operates from a 9V battery.
Size: 3.8 x 2.4 x 1 in. (97 x 61 x 25mm)
Maximum light sampling time: 80
seconds.
Quick Start
The SQM is very simple to use. Point the
SQM so that sensor/faceplate points
toward the zenith. Press the red button
once and release. Under urban skies, a
reading will be displayed almost
immediately. Under the very darkest
conditions (no moon in the sky, far from
civilization) the SQM may take up to a
minute to complete its measurement.
Please ensure that you maintain the
orientation of the SQM until the reading is
displayed.
The SQM's reading is indicative of the sky
brightness with a cone of half-angle 40
degrees centered on the faceplate
perpendicular. There must be no direct
illumination of the faceplate sensor by a
terrestrial light source if the reading is to
be meaningful.
Typical Readings
Magnitudes per square arcsecond is a
logarithmic measurement. Therefore large
changes in sky brightness correspond to
relatively small numerical changes. A
difference of 1 magnitude is defined to be
a factor of (100)
(1/5)
in received photons.
Therefore a sky brightness 5.0
mag/arcsec² fainter corresponds to a
reduction in photon arrival rate of a factor
of 100.
The following schematic gives a rough
idea of of how to interpret the readings:
At the darkest sites, natural variations in
conditions such as airglow and the
brightness of the zodiacal light are
limiting factors.
Temperature reading
The temperature in °C then °F are
displayed when you press and hold the
button a second time. Also, the model and
serial number are displayed after the
temperature.
Care of your SQM
The SQM is a fairly simple and robust
device. Avoid dropping, immersing, and
compressing it and it will give you years
of dependable service. Keep the faceplate
clean and ensure that the battery still has
useful capacity. If you have left your SQM
for a long period of time (i.e. years) and
see a white, powdery substance around
one of the battery contacts, your battery
will need to be replaced and the contacts
cleaned before you can expect reliable
operation.
The SQM should not be negatively
affected by dew during normal operation
EXCEPT for the reduction in received
light by the sensor. Make sure that the
sensor faceplate has been wiped before
making measurements.
During storage, make sure that the push-
button is not being continuously pressed
since the meter will draw current from the
battery and drain it in that situation.
Question and Answer IconNeed help?

Do you have a question about the Unihedron SQM and is the answer not in the manual?

Unihedron SQM Specifications

General IconGeneral
ManufacturerUnihedron
CategoryMeasuring Instruments
Measurement TypeNight Sky Brightness
Unitsmag/arcsec²
Accuracy±0.1 mag/arcsec²
Power Source9V battery
ModelSQM
Spectral ResponseSimilar to human eye sensitivity
Field of View20 degrees
Measurement Range0 to 22 mag/arcsec²
Resolution0.01 mag/arcsec^2
DisplayLCD

Summary

SQM Features and Operation

SQM Features

Highlights the SQM's sensitivity, temperature compensation, battery protection, calibration, precision, and display features.

Quick Start Guide

Provides instructions for initial use, including pointing the device and completing a measurement.

Typical Readings Explained

Explains sky brightness measurement scales and provides a visual guide for interpreting readings.

Temperature Reading

Details how to access and interpret temperature readings from the SQM.

Care and Maintenance

Offers advice on protecting the SQM from damage and ensuring reliable operation.

Troubleshooting SQM Issues

No Reading Displayed

Addresses issues where the SQM fails to show a reading, suggesting battery or connection checks.

SQM Auto Shut-off

Clarifies that the SQM automatically powers off to conserve battery life.

Inconsistent or Unexpected Readings

Covers reasons for readings not repeating or differing from expectations, like orientation or stray light.

Daytime or Saturated Readings

Explains why the SQM displays saturation symbols during the day or in bright conditions.

First Reading Variability

Notes that the initial reading may differ due to temperature changes upon powering up.

Additional Information and Support

Conversion Formula

Provides the formula to convert SQM readings from mag/arcsec² to cd/m².

Contacting Unihedron

Offers contact details for unanswered questions, updates, and support.

Warranty Information

States the product warranty period and mentions support for the International Dark-Sky Association.