User Guide
Univox® FSM Basic
AFILS Field Strength Meter
Part no 401014
Question and Answer IconNeed help?

Do you have a question about the Univox FSM Basic and is the answer not in the manual?

Summary

Introduction

Device Overview and Controls

Controls and Indicators Explained

Details the various controls, indicators, and range settings of the Univox FSM Basic meter.

Usage Instructions and Procedures

Background Noise Measurement

Step-by-step guide for measuring background noise levels using the meter.

Field Strength Deviation and Overspill

Instructions for measuring field strength variation and overspill for coverage assessment.

Basic Frequency Test (FQ)

Field Strength Adjustment and System Commissioning

Adjustment of Field Strength Level

Steps for calibrating the meter's field strength reading to the standard 0dB (400mA/m).

System Commissioning Procedure

Guide for integrating and verifying the AFILS system's performance and sound quality.

Safety, Care, and Maintenance

Calibration and Battery Management

Details calibration requirements and instructions for battery replacement and handling.

Technical Specifications and Environmental Disposal

Technical Specifications

Lists key technical parameters of the Univox FSM Basic meter, including power, ranges, and filters.

Environmental Disposal Guidelines

Advises on adhering to local regulations for the responsible disposal of the instrument.

Summary

Introduction

Device Overview and Controls

Controls and Indicators Explained

Details the various controls, indicators, and range settings of the Univox FSM Basic meter.

Usage Instructions and Procedures

Background Noise Measurement

Step-by-step guide for measuring background noise levels using the meter.

Field Strength Deviation and Overspill

Instructions for measuring field strength variation and overspill for coverage assessment.

Basic Frequency Test (FQ)

Field Strength Adjustment and System Commissioning

Adjustment of Field Strength Level

Steps for calibrating the meter's field strength reading to the standard 0dB (400mA/m).

System Commissioning Procedure

Guide for integrating and verifying the AFILS system's performance and sound quality.

Safety, Care, and Maintenance

Calibration and Battery Management

Details calibration requirements and instructions for battery replacement and handling.

Technical Specifications and Environmental Disposal

Technical Specifications

Lists key technical parameters of the Univox FSM Basic meter, including power, ranges, and filters.

Environmental Disposal Guidelines

Advises on adhering to local regulations for the responsible disposal of the instrument.

Univox FSM Basic Specifications

Power IconPower
power_source1 x PP3 9V alkaline
current_drain20-50mA
battery_life>10 hours
Audio Specifications IconAudio Specifications
indicating_ranges-51dB to -24dB and -24dB to +3dB
resolution3dB
absolute_accuracy±0.5dB@0dB (400mA/m)
Filter Characteristics IconFilter Characteristics
norm_filter_characteristics-3dB@630Hz & 5kHz
flat_filter_characteristics-3dB@40Hz & 14Hz
headphone_monitor_powerMax 2x19mW into 32ohms