Do you have a question about the Univox FSM Basic and is the answer not in the manual?
Details the various controls, indicators, and range settings of the Univox FSM Basic meter.
Step-by-step guide for measuring background noise levels using the meter.
Instructions for measuring field strength variation and overspill for coverage assessment.
Steps for calibrating the meter's field strength reading to the standard 0dB (400mA/m).
Guide for integrating and verifying the AFILS system's performance and sound quality.
Details calibration requirements and instructions for battery replacement and handling.
Lists key technical parameters of the Univox FSM Basic meter, including power, ranges, and filters.
Advises on adhering to local regulations for the responsible disposal of the instrument.
Details the various controls, indicators, and range settings of the Univox FSM Basic meter.
Step-by-step guide for measuring background noise levels using the meter.
Instructions for measuring field strength variation and overspill for coverage assessment.
Steps for calibrating the meter's field strength reading to the standard 0dB (400mA/m).
Guide for integrating and verifying the AFILS system's performance and sound quality.
Details calibration requirements and instructions for battery replacement and handling.
Lists key technical parameters of the Univox FSM Basic meter, including power, ranges, and filters.
Advises on adhering to local regulations for the responsible disposal of the instrument.
| power_source | 1 x PP3 9V alkaline |
|---|---|
| current_drain | 20-50mA |
| battery_life | >10 hours |
| indicating_ranges | -51dB to -24dB and -24dB to +3dB |
|---|---|
| resolution | 3dB |
| absolute_accuracy | ±0.5dB@0dB (400mA/m) |
| norm_filter_characteristics | -3dB@630Hz & 5kHz |
|---|---|
| flat_filter_characteristics | -3dB@40Hz & 14Hz |
| headphone_monitor_power | Max 2x19mW into 32ohms |