DP-10
Air Master
Operation Manual
Question and Answer IconNeed help?

Do you have a question about the URIT Medical Electronic VRN DP-10 and is the answer not in the manual?

Overview

This document is an operation manual for the VRN DP-10 Air Master, a device manufactured by URIT Medical Electronic Co., Ltd. The manual provides detailed information on the device's introduction, technical parameters, operation environment, installation, contraindications, precautions, cleaning and sterilization, maintenance, storage and transportation, accessories list, troubleshooting, after-sale service, symbols, manufacturer's rights, specification sheet of the seals, and information about powders.

Function Description

The DP-10 Air Master is intended for removing tartar and plaque from the surface of teeth, cleaning periodontal pockets, and polishing the tooth surface. It consists of an Air Master, a polisher nozzle, a handpiece, and polisher containers and joints.

Important Technical Specifications

  • Product Name: Air Master
  • Product Model: DP-10
  • Product Service Life: 5 years
  • Sound Pressure Level: The A-weighting sound pressure level produced by the Air Master does not exceed 80 dB.
  • Input Water Pressure: The required minimum value is 0.15 MPa.
  • Water Output: The water output is not less than 20 mL/min.
  • Handpiece Sterilization: The handpiece can withstand 250 cycles of sterilization without corrosion and damage.
  • Polisher Nozzle: Rotatable.
  • Water Supply Pressure (Normal Operation): 0.15 MPa to 0.4 MPa.
  • Working Pressure (Normal Operation): 0.25 MPa to 0.4 MPa. There should be no leakage of water or air within this range.
  • Water and Air Supply Connector: B-type connector according to the standard YY / T_0514-2018 dental handpiece hose connector, without optical fiber holes. This connector is included in the accessory.
  • Ambient Temperature (Operation): 5 ℃ to 40 ℃.
  • Relative Humidity (Operation): ≤80%.
  • Atmospheric Pressure (Operation): 70 kPa to 106 kPa.
  • Sterilization Temperature (Autoclave): Up to 135 ℃.
  • Sterilization Pressure (Autoclave): 0.22 MPa.
  • Sterilization Time (Autoclave): 3 to 15 minutes.
  • Sterilizable Parts: Can be sterilized at least 250 times.

Usage Features

Installation

The device consists of a nozzle, handpiece, device body, polisher container, cap, and connector.

  1. Handpiece Installation: Connect the handpiece to the device body by pushing it in the direction of the arrow.
  2. Filling the Powder: Unscrew the cap of the container counterclockwise. Fill the powder into the container. Install the cap clockwise.
  3. Subgingival Nozzle Installation: Match the subgingival polisher handpiece.
  4. Connection of Polisher Handpiece and Dental Scaler: Connect the polisher handpiece to the dental scaler.

Contraindication

Patients suffering from chronic bronchitis or asthma must not be treated with the Air Master. The jet of air and powder could cause respiratory difficulties.

Precautions

  • Read and understand the instructions before operating.
  • Follow all warnings and instructions on the device when operating.
  • The product is specially used for oral therapy and is not available for other use.
  • The product is limited to professional dentists.
  • Before each use, operate the device outside the patient's mouth for inspection. If any abnormalities such as loosening, noise, air or liquid leakage, please stop using it immediately and contact the original dealer.
  • To prevent serious damage, avoid hitting or accidentally dropping the Air Master.
  • The pressure of the water and air input must meet the requirements of this manual. If the pressure is too high, it will cause the risk of air or water leakage and bursting of the Air Master.
  • Use dry and clean compressed air, and ensure air compressor, water, and air filtration systems are maintained regularly to ensure the quality of compressed air and water. Undried compressed air will cause powder blocking in the container or handpiece.
  • After each use, pour out the remaining powder in the container, and use compressed air to clean the outer surface of the handpiece and the powder in the container. Finally, use Air Master normally without powder for 5 to 8 seconds to clean the internal tube.
  • After each use and clean, the cap of the container should be installed for storage to prevent moisture from entering the container, causing powder blocking when the next use.

Information about Powders

  1. Patients on a low salt diet must not be treated with powder containing sodium bicarbonate. For patients on a law salt diet, use the powder without sodium bicarbonate provided by VRN.
  2. The aromatic powder provided by VRN can lead to allergic reactions to the flavor in specific cases. For patients known to suffer such reactions, use the powder without flavor provided by VRN.
  3. Powder containing sodium bicarbonate provided by VRN can be used only for supra-gingival application. For sub-gingival application, use the specific powder provided by VRN and refer to its operation instructions.
  4. Never use an abrasive powder, e.g., alumina-based, as this would damage the unit.
  5. Do not direct the jet of powder toward fillings, crowns, or bridgework as this could damage these restorations.

Maintenance Features

Cleaning and Sterilization

  1. Alcohol Sterilization: Use 70%~80% (volume ratio) ethanol to soak a piece of clean dry gauze. Wipe the surface to be sterilized twice for 3 minutes. Dry or wipe the residual alcohol with a clean, dry cloth.
  2. Autoclave Sterilization: Sterilize at up to 135 ℃ and 0.22 MPa pressure for 3 to 15 minutes by an autoclave. After sterilization, the handpiece should be naturally cooled before use again to avoid scalding.
  3. Handpiece Sterilization: The handpiece can be sterilized by high temperature and high pressure. The device body and dental quick connector cannot be sterilized by high temperature and pressure.
  4. Pre-sterilization: Use compressed air to remove the cleaning fluid remaining in the handpiece.
  5. During Sterilization: Ensure to remove the tip from the handpiece and avoid mixing it with other instruments for sterilization.
  6. Forbidden Sterilization Methods: It is strictly forbidden to sterilize the handpiece in the following ways:
    • Put the handpiece in the solution and cook.
    • Soak the handpiece with disinfectant water such as iodine, alcohol, glutaraldehyde, etc.
    • Put it into the oven or microwave oven to bake at high temperature.

Maintenance

  1. Waterproof O-rings (Device Body and Handpiece): There are two waterproof O-rings at the connecting end of the device body and the handpiece. As they need to be plugged and unplugged repeatedly, once they are damaged or excessively worn, the waterproof O-rings should be replaced in time.
  2. Sealing O-ring (Device Body and Container): There is a sealing O-ring at the connecting end of the device body and container. As the container cap needs to be removed repeatedly, once it is damaged or excessively worn, the sealing O-ring should be replaced in time.
  3. Post-use Cleaning: After each use, pour out the remaining powder in the container, and use compressed air or a clean, dry cloth to clean the outer surface of the handpiece, container, or threads of the container cap, O-ring, and the residual powder in the container. Finally, use Air Master for 5 to 8 seconds without powder to clean the internal tube.
  4. Maintenance of the Product: The maintenance of the product should be done by professionals.

Storage and Transportation

  1. Storage Temperature: -20 ℃ to 55 ℃.
  2. Relative Humidity (Storage): ≤ 80%.
  3. Atmospheric Pressure (Storage): 70 kPa to 106 kPa.
  4. Installation/Storage Location: The device should be handled with care and should be installed or stored in a cool, dry, and ventilated place.
  5. Mixing Restrictions: Do not mix with toxic, corrosive, flammable, and explosive materials during storage.
  6. Earthquake Source: The device should be far away from the earthquake source.
  7. Dangerous Goods: Transport should not be mixed with dangerous goods.
  8. Shock and Vibration: Avoid excessive shock and vibration during transportation.
  9. Environmental Protection: Keep away from rain, sunlight, or snow during transportation.

Troubleshooting

  • Air leakage of container:
    • Possible cause: O-ring is aging.
    • Solution: Replace the O-ring.
  • Liquid leakage of handpiece:
    • Possible cause: O-ring is aging.
    • Solution: Replace the O-ring.
  • Air or liquid leakage of handpiece:
    • Possible cause: O-ring or seals are aging.
    • Solution: Replace the O-ring or seals.
    • Possible cause: There is no air or water.
    • Solution: Check if there is air or water in the dental quick connector.
  • When operating, powder, air or water do not escape from the polisher nozzle:
    • Possible cause: Handpiece is blocked by powder.
    • Solution: Pull out the handpiece and check if air or powder is sprayed out from the device. If powder, air, and water can be sprayed out, please use a steel wire to dredge the handpiece, and then clean the handpiece with compressed air.
  • No powder, with air supply:
    • Possible cause: There is too little powder in the container.
    • Solution: Check if there is too little powder in the container.
    • Possible cause: The supply of air is wet and the powder is caking in the container.
    • Solution: Empty and dry the container, refill the powder, and ensure a dry and clean air source.

After-sale Service

This device is guaranteed by the warranty card from the date of sale and is responsible for lifetime maintenance. Irreparable device damage caused by non-designated dedicated maintenance personnel is not covered by the free warranty.

Symbols

  • VRN Logo: Trademark.
  • Exclamation Mark in Triangle: Caution.
  • Booklet Icon: Refer to instruction manual/booklet.
  • SN: Serial number.
  • Factory Icon: Manufacturer.
  • Calendar Icon: Date of manufacture.
  • 135°C with Waves: Sterilization at 135°C in an autoclave.
  • Hourglass: Use by date.
  • Atmospheric Pressure Icon: Atmospheric pressure limitation.
  • Humidity Percentage Icon: Humidity limitation.
  • Thermometer Icon: Temperature Limit.
  • Up Arrows: This way up.
  • Broken Glass Icon: Fragile, handle with care.
  • Umbrella with Rain: Keep away from rain.
  • MD: Medical device.