Why is there no digital pressure reading on my vacuubrand CVC 3000?
- AAllen WrightSep 7, 2025
If there is no digital pressure reading on your vacuubrand Controller, it may indicate a defective pressure transducer.
Why is there no digital pressure reading on my vacuubrand CVC 3000?
If there is no digital pressure reading on your vacuubrand Controller, it may indicate a defective pressure transducer.
Why Controller does not react when pressing keys (except ON/OFF), PC symbol is displayed?
If the vacuubrand Controller does not react when pressing keys (except ON/OFF) and the PC symbol is displayed, it means the controller is in remote operation. You can either control the controller via the interface or switch off remote operation.
Why function is not displayed, no menu on vacuubrand CVC 3000 Controller?
If the function is not displayed and there is no menu on your vacuubrand Controller, it indicates that no controllable device (valve, VMS, NT VARIO/VARIO-B pump) is connected. To resolve this, connect the necessary devices or use the controller as a measuring device.
What does it mean if there is no digital pressure reading, warning triangle is flashing, five or seven blips on vacuubrand Controller?
If the warning triangle is flashing with five or seven blips and there's no digital pressure reading on your vacuubrand Controller: * For five blips: the external sensor might be defective or removed. Plug in the sensor, use a new one, or reconfigure without the sensor. * For seven blips: it could indicate that the internal sensor is defective.
What to do if warning triangle and pump symbol are flashing with six blips on vacuubrand CVC 3000?
If the warning triangle and pump symbol are flashing, accompanied by six blips, and you're using an NT VARIO / VARIO-B pump with a VMS, try disconnecting the VMS and starting the vacuubrand controller. Alternatively, there may be a fault with the NT VARIO / VARIO-B pump or the VMS itself, so check the pump and/or VMS or reconfigure. Also, ensure the connecting cables from the pump or VMS are securely attached.
What to do if venting valve does not react, valve symbol is displayed on vacuubrand CVC 3000?
If the venting valve on your vacuubrand Controller does not react, but the valve symbol is displayed, it may be contaminated. Try cleaning the valve to resolve this issue.
Why internal venting valve does not react, valve symbol is not displayed on vacuubrand CVC 3000 Controller?
If the internal venting valve does not react and the valve symbol is not displayed on your vacuubrand Controller, it could be because an external pressure transducer is connected and active. To fix this, select the internal pressure transducer or connect an external venting valve.
What to do if function “Pump down” control stops, arrow down is flashing on vacuubrand Controller?
If the function "Pump down" stops and the arrow down is flashing on your vacuubrand Controller, it indicates that the pressure is below the preselected minimum pressure. You can confirm by pressing START/STOP and adapt the pressure value if necessary.
How to solve function “Vac control” stops, arrow up is flashing on vacuubrand CVC 3000 Controller?
If the function "Vac control" stops and the arrow up is flashing on the vacuubrand Controller, it means the preselected maximum pressure has been exceeded. To resolve this, confirm by pressing START/STOP and adapt the pressure value if necessary.
Why is the clocks symbol flashing and control stopped on my vacuubrand Controller?
If the clocks symbol is flashing and the control is stopped on your vacuubrand Controller, it means the preselected process time has passed. To proceed, confirm by pressing the START/STOP button.
Type | Vacuum Controller |
---|---|
Display | LCD |
Weight | 2.5 kg |
Pump compatibility | Compatible with various vacuum pumps |
Control accuracy | ± 1 mbar |
Sensor compatibility | Pirani |
Interface | RS232, USB |
Control functions | Pressure control |
Power supply | 100-240 V AC |
Provides essential safety and general information for users of the product.
Explains manual conventions, symbols, icons, and definitions used throughout the document.
Details intended use, improper use, and foreseeable misuse of the controller.
Identifies authorized personnel and user permissions for operating the controller.
Outlines safety measures, protective clothing, and hazard elimination.
Covers safety regulations for service jobs and product return requirements.
Guidance on proper disposal of the controller and environmental considerations.
Describes the CVC 3000 controller's design, capabilities, and features.
Explains how the controller manages vacuum processes, its components, and specifications.
Lists and describes the selectable standard and optional operation modes.
Covers installation conditions and different mounting versions like table top, built-in, and direct.
Details the installation of the controller as a table-top or bench-top unit.
Explains how to install the controller as a built-in unit, including cut-out dimensions.
Describes mounting the controller directly at the vacuum vessel using a clamping ring.
Provides visual examples of controller applications with pumps and other equipment.
Covers electrical and vacuum connections, including specific details for mains and vacuum lines.
Guides on connecting the wall power supply, including preparing plugs.
Details how to connect the vacuum line, including precautions for hoses and process media.
Explains the VACUU·BUS® system for connecting accessories and address assignment.
Describes the physical controls on the controller, including the selection knob and control panel keys.
Explains the functions of the selection knob for navigation and value adjustment.
Details the functions of the control panel keys like On/Off, Start/Stop, and Vent.
Lists key combinations for accessing advanced menus and functions.
Explains the pressure display, bar graphic, numerical values, and units.
Describes the components of the pressure display, including title bar, operation mode, and values.
Illustrates and explains various icons used on the display for operation and active components.
Explains the audio signals used for keystrokes and error indications.
Describes the general structure and navigation of the controller's menus and submenus.
Provides instructions on navigating menus, selecting functions, and understanding action symbols.
Details the procedure for switching the controller on and off.
Guides on setting the controller's language and preferred pressure units.
Explains how to select and change between different operation modes.
Step-by-step guide to selecting an operation mode from the menu.
Describes the available operation modes like Pump down, Vac control, Auto mode, Program, and VACUULAN.
Instructions on how to start the vacuum controlling process after selecting an operation mode.
Covers managing the controller while a process is running, including venting and mode changes.
Explains the function and usage of the VENT button for system venting.
Describes how to switch between operation modes while the controller is running.
Instructions on how to access and view the pressure vs. time curve display.
Details how to return from the graphic display to the basic pressure display.
Explains how to adjust set vacuum or motor speed during operation.
Guides on fine-tuning or quickly adjusting the set vacuum value during operation.
Describes how to adjust the motor speed of a VARIO® pump during operation.
Instructions on how to stop the controller and the vacuum control process.
Explains how to adapt and optimize operation modes through specific menus.
Details parameters and settings for the Pump down operation mode.
Details parameters and settings for the Vac control operation mode.
Explains parameters for Auto mode, focusing on VARIO® pump functionality.
Describes how to load, edit, and store individual programs with vacuum and time presets.
Details settings for VACUULAN mode, optimized for VACUU·LAN networks.
Covers functions related to managing programs, including opening, editing, and storing.
Step-by-step guide to opening and changing existing programs.
Explains how to access and use the program editor to modify program steps.
Instructions on how to save edited programs under a specific program number.
Shows how program steps are displayed during active operation.
Explains the advantages and usage of the program memory.
Provides practical examples of program configurations for different applications.
A template for noting down program settings.
Covers setting controller parameters, adjusting sensors, and loading defaults.
Details selectable menu items within Configuration, including Defaults and Autostart.
Explains the Display and Sensors submenus within the Configuration menu.
For special controller parameters like VACUU·BUS address assignment or updates.
Details selectable menu items within the Function menu, like Venting and Vario init.
Explains how to assign addresses to VACUU·BUS components for proper communication.
Lists possible address configurations for various VACUU·BUS components.
Describes how to set up and perform differential pressure measurements using two sensors.
Explains how errors are indicated on the display using warning triangles and icons.
Provides a troubleshooting table for common faults, their causes, and remedies.
Details procedures for automatically and manually resetting the controller errors.
Explains how to handle errors related to external components like sensors or valves.
Provides instructions for cleaning the controller's surface, venting valve, and internal sensor.
Specific instructions for cleaning the controller's outer surface.
Steps for cleaning the internal venting valve.
Procedure for cleaning the controller's internal sensor.
Guides on readjusting the vacuum sensor at atmospheric pressure and under vacuum.
Contains technical data, product comparison, rating plate, and wetted parts information.
Lists specifications like ambient conditions, power supply, and electrical data.
Compares features of different CVC 3000 models and related products.
Details information found on the controller's rating plate.
Lists components and their corresponding wetted materials.
Describes commands for communication via RS232 interface.
Shows the pin assignment for the Sub-D panel connector.
Lists read commands compatible with CVC 2000.
Lists write commands compatible with CVC 2000.
Lists read commands specific to CVC 3000.
Lists write commands specific to CVC 3000.
Provides order numbers for the controller, spare parts, and accessories.
Details the service range, handling procedures, and contact information.
Defines technical terms used in the manual.
An alphabetical index of topics covered in the manual.
Includes EC Declaration of Conformity, US/CAN Certificate, and China RoHS 2 declaration.
Official declaration of the product's conformity with relevant EU directives.