Vapotherm precision flow Quick Reference Guide

Vapotherm precision flow
2 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
Attach High or Low Flow Vapor Transfer Cartridge to Disposable Water Path
Attach patient delivery tube to Disposable Water Path
Open door and install Disposable Water Path into docking station
Hang sterile water bag
Wipe water spike with alcohol pad and insert into sterile water bag
Allow a minimum of 200ml of water to ll into the Disposable Patient Circuit
Precision Flow is ready for start up
Press the Run/Standby button. Unit will enter Standby mode.
Clamp the water inlet tube and disconnect it by pulling out the spike.
Open the door, remove the Disposable Patient Circuit (includes delivery
tube & cartridge) by sliding it upwards out of the docking station
Discard all disposables according to hospital guidelines.
Disconnect unit from AC power.
QUICK REFERENCE GUIDE
VAPOTHERM, INC.
22 Industrial Drive
Exeter, NH 03833
T: 603-658-0011
USA
3006001 Rev. D
This guide provides you with basic instructions on how to set up and operate the
Precision Flow.™ Before operating the Vapotherm Precision Flow,™ please review the
Operators Instruction Manual which can be found at our website www.vtherm.com
Blocked tube
General fault
Water out
Alarm mute
LED
Alarm mute
button
Setting Control
Knob
Battery low or
charging (not
indicative of charge)
level)
Disposable
Water Path
faulty or absent
Low Flow Vapor
Transfer Cartridge
High Flow
Vapor Transfer
Cartridge
Vapor Transfer
Cartridge fault (Low
ow & High ow)
Gas supply fault
Run/Stop
Status LED
Run/Standby
button
QUICK REFERENCE GUIDE
Technical Support
T: +1 (603) 658-5121 International
855 557 8276 Domestic
ts@vtherm.com
Install oxygen sensor
Install gas inlet lters on back with lter bowls vertical (glass side down).
Attach air & O₂ hoses. Plug in power cord.
Rotate the blue Setting Control Knob to illuminate display
Press in Setting Control Knob to select the parameter and rotate to adjust
the value
Press and release (do not hold) the Run/Standby button once to start.
A GREEN light indicates RUN mode (AMBER light indicates STANDBY mode,
No Flow)
Green light will stop ashing once temperature is reached
The ashing green LED becomes steady when the set variables are reached.
Place the cannula on the delivery tube
After cannula is attached, it is ready to be placed on patient
To activate the internal backup battery mode, the Run/Standby LED must
be green before disconnecting the power cord.
The unit will enter into battery mode and will maintain ow and oxygen
percentage for at least 15 minutes.
The battery icon will ash.
DISPOSABLE PATIENT CIRCUIT
START UP AND ADJUSTING PARAMETERS
CONNECT TO PATIENT
BATTERY BACK-UP MODE
SHUT DOWN MODE
May be patented. www.vtherm.com/patents

Other manuals for Vapotherm precision flow

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vapotherm precision flow and is the answer not in the manual?

Vapotherm precision flow Specifications

General IconGeneral
Oxygen Concentration21% - 100%
HumidificationNear 100% relative humidity
Power Supply100 - 240 VAC, 50/60 Hz
Therapy TypeHigh flow nasal cannula (HFNC) therapy
Heater TemperatureAdjustable, typically 34°C - 37°C