Before using the micromotor, verify the
correctness of the changed parameters.
7.7. Factory Default Parameters
To return to the original default parameters,
follow the general reset instructions:
• Make sure that the battery charger is
disconnected.
• Switch off the device.
• Keep holding down the CONFIRM key ✓ and
the POWER key simultaneously. The device
will switch on and the display will read:
NOTE
Be aware that all individual settings,
including the settings in Dr`s Choice
program, will be deleted when “reset to
default parameters” has been carried out.
The audible signals settings are not affected:
they remain with the last setting used
before the device was switched off.
ASR function can be modified for each file,
but the setting will return to the default
settings as soon as a different file is
selected.
7.8. Maintenance
Routine Maintenance
• Inspect the micromotor cable at least once
every six months. If any deterioration of the
sheathing is noticed, have the cable replaced by
an authorized service center.
• Check for liquids or smoke exiting the device. In
this case, remove the power supply from the
device immediately and contact an authorized
service center.
• For optimal battery performance, replace the
battery every 2 years.
• For the maintenance, e.g. lubrication of the
VDW Endo 6:1 contra-angle, refer to the separate
contra-angle operating instructions and check the
additional warnings below.
WARNINGS
• The contra-angle should be lubricated
after cleaning and disinfecting, but prior to
sterilization.
• When lubricating the contra-angle, check
that no lubricant penetrates the micro-
motor.
• If the contra-angle is lubricated manually,
please ensure excess oil is removed by
compressed air (blow for approx. 5 sec.)
before contra-angle is being put back on
the micromotor. Calibrate after lubrication.
• If the contra-angle is lubricated auto-
matically in a maintenance or lubrication
device, please follow the instructions for
use of the device manufacturer carefully
and ensure no excess oil is left in the
processed contra-angle.
• Do not lubricate the micromotor for any
reason. Lubricant may contaminate the
micromotor causing damage and result in
unsafe operation. This will invalidate the
warranty.
• Do not introduce any foreign objects into
the micromotor shaft.
Special Maintenance
Contact VDW GmbH for any special maintenance
that may be required.
NOTE
Some operations carried out during repair
could lead to deletion of personal settings
such as changed torque, speed or Dr’s
Choice settings.