EasyManua.ls Logo

VEE GEE BX-3 - User Manual

VEE GEE BX-3
2 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Range: Brix 60.0-92.0%
Resolution: 0.2%
Accuracy: ±0.2%
Dimensions: 40 x 40 x 140mm (1.6 x 1.6 x 5.5”)
Weight: 230g (8.1 oz.)
Supplied With: Vinyl Carrying Case (1), Plastic Transfer Pipet (1)
Optional Oil Standard: Brix 60.0%, ¼ oz. Bottle (cat. no. 45000-OS025)
Printed in China
VGMNL010320-43005
Operation Manual
Model BX-3
Brix 60-92%
Specifications
Cat. No. 43005
Temperature Correction Table
For Refractometers Calibrated At 20°C (l=589nm)
Source: ICUMSA, 1
974
As an alternative to recalibration of the refractometer due to ambient temperature changes, the following temperature
correction table can be used. Ensuring that both the distilled water and the ambient temperature are both at exactly 20°C,
follow steps 2-7 under the CALIBRATION section on the preceding page. Once this is accomplished, if ambient temperature
changes, simply apply the temperature correction values listed below. For example, a reading of Brix 40.0% at 25°C is
corrected to Brix 40.4%; a reading of Brix 65.0% at 10°C is corrected to Brix 64.25%.
Warranty information a
nd
registration form can be found at:
www.veegee.com/service_support
Question and Answer IconNeed help?

Do you have a question about the VEE GEE BX-3 and is the answer not in the manual?

Summary

Temperature Correction Table

Correction Table

Table providing temperature correction values for refractometer readings based on ambient temperature.

VEE GEE Specifications

Refractometer Components

Rubber Hood

Protects the focusable lens and prevents ambient light entry during use.

Calibration Ring

Mechanism for zeroing or calibrating the refractometer without extra tools.

Non-Roll Stand

Prevents instrument damage when placed down between measurements.

Daylight Plate & Prism

Covers the prism and sample during readings.

Rubber Grip

Provides insulation against hand heat for accurate results.

Calibration Ring Set Screw

Secures the calibration ring to prevent shifts during use.

Calibration Procedure

Apply Liquid Standard

Apply one or two drops of liquid standard onto the prism surface for calibration.

Close Daylight Plate

Close the daylight plate, ensuring the liquid standard covers the entire prism surface evenly.

Adjust Focus

Adjust the eyepiece to bring the scale into focus by turning it clockwise or counterclockwise.

Verify Calibration Point

Confirm the boundary line crosses the scale at the correct calibration point (e.g., '85').

Adjust Calibration Ring

Loosen the calibration ring and adjust it until the boundary line is at the correct calibration point.

Final Wipe

Gently wipe the prism with tissue paper after calibration is complete.

Precautions

General Use

Apply Sample Solution

Apply one or two drops of sample solution to the prism surface, holding it at an angle.

Close Daylight Plate

Gently close the daylight plate, ensuring the sample covers the prism surface evenly.

Adjust Focus

Adjust the eyepiece to bring the scale into focus by turning it clockwise or counterclockwise.

Read Brix Value

Read the Brix% value where the boundary line of blue and white fields crosses the scale.

Clean Prism

Clean the prism using tissue paper and water after each measurement.

Overview

The VEE GEE Refractometer, specifically the Model BX-3, is an optical instrument designed for measuring the Brix% of solutions. It is built with modern designs to provide reliable performance over many years, provided it is handled with proper care. The device is calibrated at 20°C (λ=589nm) and includes a temperature correction table to account for ambient temperature changes, eliminating the need for frequent recalibration.

Function Description

The primary function of the VEE GEE Refractometer is to determine the Brix% of a liquid sample. Brix% is a measure of the sugar content in a solution, and the device provides a direct reading on an internal scale. The refractometer operates by measuring the refractive index of a liquid sample, which correlates to its concentration.

The instrument features a Rubber Hood that houses and protects the focusable lens, preventing external light from interfering with readings through the eyepiece. The Eyepiece allows the user to view the internal scale and the boundary line between the blue and white fields, which indicates the Brix% reading.

A Calibration Ring is integrated into the design, allowing for easy zeroing or calibration of the refractometer without the need for additional tools. This ring can be turned clockwise or counterclockwise to adjust the boundary line during calibration. A Calibration Ring Set Screw secures the calibration ring in place, protecting it from accidental shifts.

The Rubber Grip insulates the refractometer from hand heat, ensuring accurate results by preventing temperature fluctuations from the user's hand from affecting the measurement.

The Daylight Plate & Prism assembly is where the sample is placed. The prism is made of optical glass and is covered by the daylight plate during readings. The daylight plate is opened to apply the sample and then closed to create a thin, even layer of the liquid over the prism surface.

A Non-Roll Stand is included to protect the instrument from damage when it is set down between measurements.

Usage Features

Before using the refractometer for general measurements, it must be calibrated. Calibration should be performed at the start of each day or whenever significant shifts in ambient temperature occur. The standard calibration procedure involves opening the daylight plate and applying one or two drops of a liquid standard (e.g., a Brix 85.0% oil standard or a prepared sucrose solution of known concentration, minimum Brix 60.0%) onto the prism surface. It's crucial that the liquid standard covers the entire prism surface without bubbles or gaps; if not, it must be reapplied.

Once the standard is applied, the daylight plate is gently closed. The user then looks through the eyepiece, directing the daylight plate towards a light source. If the scale is out of focus, the eyepiece (rubber hood) can be gently turned clockwise or counterclockwise to bring it into focus. The internal view will show an upper blue field and a lower white field, with a boundary line separating them. For calibration, this boundary line must align with the known value of the liquid standard (e.g., "85" for an 85.0% Brix standard). If the boundary line is not aligned, the set screw on the calibration ring is gently loosened, and the calibration ring is turned until the boundary line aligns correctly. After alignment, the set screw is gently tightened.

As an alternative to recalibration due to ambient temperature changes, a temperature correction table is provided. This table allows users to adjust their readings based on the ambient temperature and the measured Brix%. For example, if a reading of Brix 40.0% is taken at 25°C, the corrected value would be Brix 40.4%. Similarly, a Brix 65.0% reading at 10°C would be corrected to Brix 64.25%. This feature ensures accuracy even when the ambient temperature deviates from the 20°C calibration temperature.

For general use, the process is similar to calibration. The daylight plate is opened, and one or two drops of the sample solution are applied to the prism surface. The plate is then gently closed, ensuring the sample spreads evenly across the prism. The user looks through the eyepiece, directing the daylight plate towards light, and adjusts the focus if necessary. The reading is taken at the point where the boundary line between the blue and white fields crosses the scale. This value represents the Brix% reading of the sample.

Maintenance Features

Proper care and maintenance are essential for the longevity and accurate performance of the refractometer.

The instrument is an optical device and should be handled gently to prevent damage from drops or rough handling.

The prism, being made of optical glass, is susceptible to scratches. Therefore, no rough or abrasive materials should be used for cleaning the prism.

After each use, the prism surface and daylight plate must be cleaned. This is done by wiping them with a soft cloth or tissue soaked in water, followed by wiping with a dry cloth or tissue.

It is important not to hold the refractometer under a stream of water from a faucet, nor should it be splashed with or dipped in water.

If the prism surface becomes coated with an oily solution or similar residue, it may repel test samples and lead to inaccurate readings. In such cases, the prism should be cleaned with a weakened detergent or a similar solvent.

These maintenance practices ensure the refractometer remains in optimal condition for consistent and reliable measurements.

VEE GEE BX-3 Specifications

General IconGeneral
ScaleBrix
Brix Range0-32%
Accuracy±0.2%
DisplayLCD
ATC10-30°C
Measurement Range0-32% Brix
Resolution0.2% Brix
Power Supply2 x AAA batteries

Related product manuals