5000 Enter the value of the known test weight as described above.
07 SEC The display will count down insuring a stable reading.
CAL d The meter has accepted the calibration parameters. Press PRINT key
to advance to final step.
FCd 0 Press PRINT to store the new settings in non-volatile memory.
StorE
donE The meter now reboots to make use of the new constants.
5000 The display shows the value of the weight applied and is ready for
normal operation.
Note: Selecting 1 for FCd will exit the calibration mode without storing the new values. Selecting 2 for FCd
will store pre-programmed factory default values.