Powering Off the Cellcorder
1.
Press and hold the green Power Key.
2. Powering off doesn’t automatically delete
voltage or resistance readings from memory.
3. If readings are in memory, you are prompted
to save them.
Power
Taking Voltage Readings
1.
Connect the voltage probes to the Cellcorder
and select Test Mode/F1.
2. From the Test Mode menu, select
F1-Voltage.
3. If "Resume Test" appears, put the probes
on the cell.
4. If "Stored in Memory" appears, retest the
cell or change the cell number by typing a
number and pressing Enter.
5. When the unit beeps, remove the probes.
The cell number advances automatically
to the next cell.
Don't measure voltages greater than 20V DC.
Warning