J
James McculloughAug 3, 2025
What to do if VET-PRO VIP 2000 shows downstream occlusion?
- SScott GuerreroAug 3, 2025
If your VET-PRO Medical Equipment indicates a downstream occlusion, several factors could be the cause. Check if the roller clamp is closed and open it. Ensure the set is not kinked; straighten it out if necessary. Verify that the administration set is loaded correctly; reload it according to the instructions. The cannula might be blocked, so consider changing it. Low temperature can affect tubing elasticity, so warm the liquid and bring the tubing to room temperature (above 19°C). You may also need to adjust the pressure setting.