VET-PRO Operator’s Manual
33
To continue the interrupted program, press the
START/OK key.
To repeat the previous program or start a new
program
, press the
STOP/NO
key.
Volume 1256 ml
Max Rate: 234 ml/h
T
P
N
Press OK to start
If you choose to complete an interrupted
program, the pump will display interrupted
program data. Scroll with the ARROW keys to
check program data. Pressing the START/OK
key will start the program operation.
Program Setting
F
Note:
Whenever you decide to set a new program, follow the
instructions at the bottom of the screen.
Note:
-If one of the values is changed during the programing mode,
the pump will not store the new value, unless the change is
confirmed by pressing the START/OK key.
-If the flush key is activated during data setting, the pump
will return to the original (previous) screen, once flushing
has been completed.
Volume 1256 ml
1.
The first parameter to set is the volume of the
TPN bag.
T
P
N
Press OK key
Total Time 12:30
T
P
N
Press OK key
2.
Set the total time for the program to run:
Options are:
a. Setting 4 digits - 2 first digits will represent
hours and the other 2 will represent minutes
b. Setting 2 digits for full hours
c. Setting 1 digit for hours and using the digital
point to set minutes.
d. Using the digital point to set only minutes.
Up Time: 0:30
3.
Set the time for the rate to taper up
(up to 4:15 h).
T
P
N
Press OK key
Down Time: 0:30
4.
Set the time for the rate to taper down
(up to 4:15 h).
T
P
N
Press OK key