How to fix Generator board A10 LED H1 ON in Villa Sistemi Medicali Medical Equipment?
- RRhonda HopkinsAug 15, 2025
If the Generator board A10 LED H1 is ON, check fuses F1 and F2 on the Generator board. If they are not blown, check the line voltage to the board and to the unit and replace the Generator board if the line voltage is present and OK. If the fuses are blown, replace them with new ones respecting the value and types; perform a new exposure and check if they blow again. If they blow, check, on the Generator board, if V6 and/or V31 and V36 are short-circuited and in this case replace the tubehead and Generator board. If not short circuited, replace the tubehead.