Table of Contents

Questions and Answers:

Villa Sistemi Medicali STRATO 2000 Digital Specifications

General IconGeneral
Voltage Range40 - 150 kV
Focal Spot Size0.6 / 1.2 mm
X-ray TubeRotating Anode
Digital Detector TypeFlat Panel Detector
Exposure Time Range1 ms - 10 s
Image ReceptorDigital Detector

Summary

Safety Information and Warnings

Device Usage Warnings

Details critical warnings related to device operation, environment, and maintenance.

Installation and Service Safety Precautions

Emphasizes safety during installation and service, especially regarding high voltage.

Product Description and Components

Service Program Access and Functions

Explains the service programs accessible via passwords for unit configuration and testing.

Hardware Configuration Passwords

Details specific passwords for configuring the unit's hardware settings.

Technical Specifications and Data

Safety and Compliance Standards

Lists the safety standards and directives the STRATO 2000 Digital complies with.

Measuring Technical Factors

Details procedures for measuring technical factors like kV and mA.

Warning: High Voltage During Measurements

Warns about high voltage present when accessing internal parts for measurements.

Verifying Exposure Parameters

Explains the non-invasive method for verifying exposure parameters.

Pre-Installation Requirements

Electrical Setup and Connections

Details electrical supply requirements, connection procedures, and safety devices.

Installation Procedures

Panoramic Alignment Verification

Details the procedure for checking and adjusting the panoramic alignment using lasers.

Cephalometric Alignment Verification

Guides the user through the cephalometric alignment check using a PC and software.

Maintenance and Servicing Procedures

Technical Service Department Maintenance

Details annual inspection activities performed by authorized technical personnel.

Verification and Centering Adjustments

Covers necessary verifications and adjustments for centering the device's components.

Panoramic Axes Alignment

Describes the procedure for aligning the axes for panoramic examinations.

Laser-Based Axes Centering

Details centering the rotation, X, and Y axes using laser alignment.

Panoramic X-ray Beam Centering

Explains how to center the X-ray beam for panoramic imaging after laser alignment.

Cephalometric Axes Alignment

Outlines the procedure for aligning axes for cephalometric examinations.

Cephalometric Rotation Arm Alignment

Details the alignment of the rotation arm for cephalometric functions.

Cephalometric Ear Centering Circle Setting

Describes how to set the ear centering circles for cephalometric alignment.

Cephalometric Secondary Collimator Centering

Guides the centering of the secondary collimator for cephalometric mode.

Cephalometric Sensor Centering

Explains the procedure for centering the cephalometric sensor.

Cephalometric Soft Tissue Filter Adjustment

Details the adjustment procedure for the Soft Tissue Filter in cephalometric mode.

Rotation Arm Alignment Procedure

Details the steps for aligning the rotation arm using lasers and software.

Ear Centering Circles Calibration

Explains how to set the ear centering circles using specific software and tests.

Secondary Collimator Centering Procedure

Guides the centering of the secondary collimator using software and X-rays.

Ceph Sensor Centering Procedure

Details the centering process for the cephalometric sensor using software.

Soft Tissue Filter (STF) Adjustment

Explains how to adjust the Soft Tissue Filter using software and X-ray exposure.

Verifying kV and Exposure Time

Describes the procedure to measure kV and exposure time using invasive methods.

kVp Measurement Procedure

Provides detailed steps for measuring kVp using a multimeter and test points.

Exposure Time Verification

Explains how to verify exposure time accuracy using an oscilloscope.

mA Measurement Procedure

Details the invasive method for measuring mA values on the HF generator board.

Troubleshooting Guide

Interpreting Displayed Messages

Explains the different types of messages displayed by the unit for operational and error conditions.

Troubleshooting: Motor Movement Errors

Lists error codes related to motor movements and their corresponding reference chapters.

Troubleshooting: Primary Collimator Errors

Lists error codes for the primary collimator and their reference chapters.

Troubleshooting: Soft Tissue Filter Errors

Lists error codes for the soft tissue filter and their reference chapters.

Troubleshooting: Secondary Collimator Errors

Lists error codes related to the secondary collimator and their reference chapters.

Troubleshooting: CEPH Sensor Errors

Lists error codes for the CEPH sensor and their reference chapters.

Troubleshooting: HF Board and Tubehead Errors

Lists error codes related to the HF board and tubehead, with reference chapters.

Troubleshooting: Button Errors

Lists error codes related to button functionality and their reference chapters.

Troubleshooting: Memory Checksum Errors

Lists error codes for memory checksum issues and their reference chapters.

Service Program Descriptions and Access

Provides an overview of service programs, their levels, and access methods.

Service Password 94: System Component Testing

Describes how to use Password 94 for testing various system components.

Input Ports Test Procedure

Explains how to test the CPU board's input signals and microswitch status.

Keyboard Functionality Test

Details how to test the functionality of the keyboard keys and indicator lights.

Error E560: Button Press During Startup

Explains how to diagnose and resolve error E560 related to button presses during startup.

Error E561: X-ray Button Release During Emission

Details troubleshooting for error E561 when the X-ray button is released during emission.

Error E562: X-ray Button Pressed During Startup/Prep

Explains how to diagnose and resolve error E562 related to X-ray button presses during startup or prep.

Error E563: Remote X-ray Button Pressed During Startup/Prep

Details troubleshooting for error E563 concerning the remote X-ray button.

kV/mA Piloting Circuits Test

Describes a test to verify CPU and HF board connectivity and piloting signals.

Error E550: No Voltage on HF Board

Guides troubleshooting for error E550 indicating no voltage supply to the HF board.

Error E551: Overvoltage Condition

Details troubleshooting for error E551 related to overvoltage on the HF board.

Error E552: Overload Condition

Explains troubleshooting for error E552 indicating an overload condition.

Error E553: Broken Filament Detected

Details troubleshooting for error E553 when a fault is detected in the X-ray tube filament circuit.

Error E554: No X-ray Output

Guides troubleshooting for error E554 when the system does not sense X-ray ON signal.

Error E555: X-ray Output Too Long

Details troubleshooting for error E555 indicating an unwanted X-ray emission.

Service Password 118: Motor and Sensor Calibration