DESCRIPTION: the projection shows «Lo.5» and the
value alternately.
PROBLEM: - reading with “face”
button: the room
temperature is between 10,0 and 15,9°C (50.0/60.6ºF).
- reading with “home”
button: the room temperature
is between 5 and 9,9°C (41/48.2ºF).
SOLUTION: the temperature reading can be taken but
accuracy is not guaranteed.
DESCRIPTION: the projection shows “Lo.5”.
PROBLEM: the room temperature is too low (below 10,0°C/50.0°F if you’ve
pressed the “face”
button or below 5,0°C/41.0°F if you’ve pressed the
“home”
button).
SOLUTION: move to another, warmer room and, if you are taking the body
temperature, wait for the stabilization of the device and of the subject.
Messages about the detected temperature
DESCRIPTION: the projection shows «Hi.2» and the
value alternately when you press the “face”
button.
MEANING: ATTENTION! The temperature is
40,0°C/104.0°F or more.
DESCRIPTION: the projection reads “Hi.2”.
PROBLEM: - reading with “face”
button: the temperature detected
exceeds the limit for operation with this button (>43,0°C / >109.4ºF).
- reading with “home”
button: the temperature detected exceeds the
limit for unit operation (>80,0°C / >176.0°F).
SOLUTION: - reading with “face”
button: please make certain that
you’ve pressed the correct button, with regard to the measured object, and
that the warnings have been met.
- Reading with “home”
button: the temperature reading cannot be taken.
DESCRIPTION: the projection shows “Lo.3”.
PROBLEM: - reading with “face”
button: the forehead temperature ap-
pears to be too low (<34,0°C / <93.2°F).
-
reading with “home”
button: the surface temperature read is below the
operating limit (<1.0°C / <33.8°F).
SOLUTION: - reading with “face”
button: make certain the sensor is
not soiled or damaged and that the subject has not come from a cold room.
- reading with “home”
button: the temperature reading cannot be taken.