16
Other readings
Measurement range: 1,0/80,0°C (33.8/176.0°F)
Accuracy:
from 1,0 to 33,9°C: ±0,3°C from 33.8 to 93.1°F: ±1.8°F
from 34,0 to 35,9°C: ±0,3°C from 93.2 to 96.7°F: ±0,5°F
from 36,0 to 39,0°C: ±0,2°C from 96.8 to 102.2°F: ±0,4°F
from 39,1 to 43,0°C: ±0,3°C from 102.3 to 109.4°F: ±0,5°F
from 43,1°C to 80,0°C: ±1,0°C from 109.5 to 176.0°F ±1.8°F
Ambient temperature (shown when in stand-by)
Measurement range: 5,0/45,0°C (41.0/113.0°F)
Accuracy: ±1.0°C (±1.8°F)
(1)
In rooms where the temperature is between -10 and 15,9°C (14 and 60.7°F) and when the FACE
button is used, or in rooms where the temperature is between -10 and 9,9°C (14 and 49.9°F) and when
the HOME button is used, accuracy and the operating range are not guaranteed and the message “Lo.5”
and the temperature value are displayed alternately. In rooms where the temperature is between 40,1
and 45,0°C (104.1 and 113.0°F), accuracy and the operating range are not guaranteed and the message
“Hi.4” and the temperature value are displayed alternately.
(2)
ASTM E1965-98-(2016) laboratory accuracy requirements in the display range of 37 to 39°C (98 to
102°F) for IR thermometers is ±0,2°C (±0.4 °F), whereas for mercury in-glass and electronic thermome-
ters, the requirement per ASTM Standards E 667-86 and E 1112-86 is ±0,1°C (±0.2°F).
13.2 General specifications
- Power supply: 4 AAA (LR03) alkaline batteries - 1.5 V (included)
- Life of high quality batteries: up to 3 years or 30,000 readings (depending on use)
- Dimensions: 144 x 43,5 x 21,5 mm (5.67 x 1.71 x 0.85 inches) - including cap.
- Weight: 98 gr. (3.46 oz.) - batteries included
- Distance from the subject: calculated using an optical aiming system (approx. 6
cm/2.36 inches).
- Large, clearly visible, backlit display in 5 colors: light blue, green or violet, depending
on the button used (respectively, “face”, “home” or “Mem”); orange and blue depen
-
ding on the function (respectively, room temperature – par. 5 and MQCS – par. 8)
- Expected life: 10 years.
- Atmospheric pressure range of operating conditions: from 700 hPa to 1,060 hPa.
- Relative humidity range of operating conditions: from 15% to 93%, non-condensing.
- Protection degree IP22
- Keep in a clean, dry place, preferably at a temperature ranging between +16 and