Safety instructions
B
Intended use
A
Blood pressure measurements on children require specialist knowl-
edge! Please consult your doctor if you wish to measure the blood
pressure of a child. Do not under any circumstances use the unit on
a baby/infant.
This device is intended for persons over 12 years of age for non-invasive
self-measurement of the systolic and diastolic blood pressure, determin-
ing the pulse rate and calculating the pulse pressure.
This unit uses the oscillometric method for measuring blood pressure and
pulse rate.
During pregnancy and in cases of pre-eclampsia, the device may only
be used after consulting your doctor.
The results of the automatic blood pressure monitor can be distorted
by cardiac dysrhythmia, e.g. atrial fibrillation, premature ventricular
contraction or arteriosclerosis. Only use the blood measurement de-
vice by yourself if you have discussed this with your doctor first.
The device may only be used at the wrist and not on any other ex-
tremities.
Do not under any circumstances place the cuff over a critical area, e.g.
a wound, aneurysm, etc. or on an arm with an arteriovenous shunt.
Risk of injury! Any supply via an intravascular access point (infusion)
or other medical monitoring devices could possibly be interrupted.
Do not use the unit without first consulting your doctor if you are un-
dergoing dialysis treatment or taking anticoagulants, platelet aggre-
gation inhibitors or steroids. Internal bleeding may be caused under
these conditions.
The display of the pulse frequency is not suitable for checking the fre-
quency of cardiac pacemakers. Cardiac pacemakers and blood pres-
sure monitors do not influence each other regarding their mode of
operation.
1. Important patient information
The device is not suitable for use in close proximity to electrosurgical
instruments or magnetic resonance imaging equipment.
Do not use the device in explosive environments such as in the vicinity
of flammable anaesthetics or in an oxygen chamber.