SHORT-TERM CLOSURE OF THE AIR SUPPLY
In a situation when it is necessary to transport a patient with a bed for a short distance very
quickly, you can disconnect the mattress from the pump and close the air outlet.
To do this, switch off the pump, unplug the power cord from the wall outlet.
Remove the connecting element from pump jack and close the air outlet.
After transporting the patient you should attach the mattress to the pump as soon as possible.
Open the air outlet on the connecting element.
Connect the connecting element to the outlet of the pump.
Plug the power cord into the outlet.
Switch on the pump.
PROBLEM SOLVING
Do not try to repair the pump, do not open it as it can lead to electric shock!
The knob is in "ON" position but the pump does not work.
- Please, check if the cable is plugged to electric supply and to the pump. If not, you should plug
it.
- Please, check if there is a voltage in the electric circuit. If there is not, find the cause of such
situation.
- Please, check if the cable is not damaged. If so, please contact the manufacturer.
The pump works too slowly or the pressure is too low.
- Please, check if the pressure regulator is set on an appropriate pressure. If not, please, regulate
it.
- Please, check if the pipes are correctly connected to the air outlet. If not, press them in such a
way which prevents the air from escaping.
- Please, check if the pipes are not folded or pressed. If so, please, unbend them.
- Please, check if the cells are not damaged. If so, please, repair it using the reparation kit.
- Please, check if the pump works correctly if you can hear some strange sounds like hissing
etc., please contact the manufacturer.
CARE AND MAINTENANCE
Disconnect the pump from electricity. Clean all the surfaces with a moist cloth. Before
switching on it has to be dry.
The mattress should be cleaned with water with soap. After cleaning, dry it carefully. The
mattress mustn't be washed in high temperature, dried in a washing machine, bleached,
chemically washed, ironed or pressed.
You must not use:
solvents,
toilet detergents,
sharp brushes and hard objects,
detergents that contain chlorine,
corrosive detergents,
Do not clean the exerciser with pressure or steam aggregates!