VOLTECH AT3600 USER MANUAL TEST RESULTS HANDLING
This section shows available functions for controlling an attached instrument via the
server. This is to allow basic instrument control in applications where the AT is being
used entirely remotely. The functions available allow a program to be loaded into an
AT, and give the ability to run and stop it.
Variant LoadProgramToAT (LPCSTR ComPort, LPCSTR ProgName)
“PROGRAM DOES NOT EXIST” Program is not in the servers program folder.
“INVALID PROGRAM” Program fails AT test program validation.
“LOADING” Returned when command is first sent to
server, and every subsequent time it is sent
until loading finishes.
"LOADED" The program has been successfully loaded,
this will continue to be returned, and the
program will not load until either StopAT()
is called or the server receives results on that
COM port.
Variant RunProgramInAT (LPCSTR ComPort)
“NO PROGRAM” The AT has no program to run.
"SENT" Returned when the program has started to
run and every subsequent time the function is
called until the server receives results.
“DONE” The program has finished running, results
will be available. If the function is called
after it has returned this value it will cause
the program to be run again.
Variant StopAT (LPCSTR ComPort)
“STOPPED” The AT must stop whatever test it is running
UING THE SERVER VPN 98 - 024 PAGE 4.3.27.