VOLTECH AT3600 USER MANUAL SELF TEST
2. SELF-TEST
Self-test is a sequence of checks made by the tester to ensure the unit is functioning
correctly.
You may run Self-test at any time by pressing the SELF-TEST soft-key in the Top
Level menu.
!
Before commencing self-test, please ensure that there is nothing touching any of
the test nodes on the top surface of the tester. In particular, remove all test
fixtures and transformers.
As the self-test is running, the front panel display will indicate progress using a
horizontal bar:
CAUTION:- If the connections have been correctly made to the Safety
Interlock, then a complete test will be executed, including high voltage testing up
to 7000V. These voltages will be present on all the test nodes on the top surface
of the tester.
!
FRONT PANEL OPERATION VPN 98 - 024 PAGE 6.2.1.