SELF TEST VOLTECH AT3600 USER MANUAL
PAGE 6.2.2. VPN 98 - 024 FRONT PANEL OPERATION
At the end of the self-test, you should see the following screen:
To ensure operator safety, the high voltage sources are not checked until about 10
seconds after the start of self-test.
If the safety interlocks are not ‘made’ before the end of this period, high voltages will
not be generated or checked, and at the end of the test, you will see the following
message:
Without a safety interlock system the unit cannot perform any high voltage self tests,
hence the high voltage sources are disabled.
If there is a failure at any point in the self-test, you will see a display similar to the
following:
The message shown on the lower two display lines will depend on what caused the
failure.
In the unlikely event that your tester continues to fail, please make a note of the
failure message and contact your Voltech supplier for service.