EXECUTE MODE VOLTECH AT3600 USER MANUAL
Or
Cause:
The directory set-up in the server application as the location of test programs
does not contain a test program file with the requested part name.
b) Local program store (see chapter 2, section 2.3.4)
Cause:
The requested part number is not resident in the group of programs that have
been downloaded into the AT.
At all the above displays, pressing the CONTINUE soft key will take you back
to the top level display.
2. INDICATIONS DISPLAYED WHILE TESTS ARE RUNNING
a)
Where 1 represents the position of the test within the test program.
The operator may terminate the measurements at any time by pressing the
STOP soft-key. Alternatively, you may use the STOP input signal on the
tester’s remote port, which may be connected, for example, to an emergency
stop switch.
As the measurements proceed, the test number will be incremented, and for the
Hi-Pot and Stress Watts tests, the measured test voltage will also be displayed
on the lower line.
PAGE 6.3.4. VPN 98 - 024 FRONT PANEL OPERATION