VOLTECH AT3600 USER MANUAL EXECUTE MODE
If the FAIL indicator illuminates at this point this indicates a
problem with the compensation procedure, check the fixture
and short-circuit header connection, then re-try the fixture
compensation.
Illumination of the PASS indicator represents a successful
compensation. Remove the short circuit links and continue.
Place the component to be tested in the fixture and press the
RUN soft-key, or the footswitch.
After the measurement, remove the component and place it in
the ‘Pass’ or ‘Fail’ bin as appropriate.
Continue in this way until each component in the batch has
been tested, then press FINISH.
NOTES:
1. ERROR MESSAGES WHEN ENTERING THE PART NUMBER
a) Server program store (see section 6.2 in this chapter)
Possible causes:
Faulty Cable: The cable from the AT server port to the server PC has a
broken wire, or is not plugged in correctly.
COM-Port: The COM Port has not been set up in the server application.
FRONT PANEL OPERATION VPN 98 - 024 PAGE 6.3.3.
Server not running: Either the server PC is not powered on, or the server
application is not running.