MEASURE MODE VOLTECH AT3600 USER MANUAL
pressing ENTER takes you to the relevant set up display:
The details of the set up display will vary depending on which test you have selected. For
the capacitance test, the list of parameters to enter is: Test voltage, Test frequency,
Integration, Hi node(s), and Lo node(s). For Winding Resistance, the list is simply:
Integration, Hi node, and Lo node.
Some tests, such as Winding Inductance and Capacitance, will require you to set up the
measurement conditions - the test voltage and frequency. If these are known, you can enter
the values you require.
Alternatively, the tester allows a choice of AUTO for the measurement conditions for some
tests. Choose this if you are unsure of what voltage and / or frequency to program.
Note that there are sometimes restrictions on which combinations of parameters can be
AUTO and SPECIFIED. In the ‘C Capacitance’ test, the valid combinations are as follows:
Voltage
Auto Auto Specified
Frequency
Auto Specified Specified
In the set up list, it is always the parameter on the third line that is active, for example ‘Test
voltage’ in the display above.
Parameter data is either typed in via the keyboard (if the data to be entered is for example a
number of volts), or selected by pressing the third soft-key. Examples of using the third
soft-key are: selecting between AUTO and SPECIFIED for the ‘Test voltage’; and selecting
the ‘Integration’:
In this particular case, the Integration is set to MEDIUM, and the next available choice is
LONG. Pressing the third soft-key repeatedly will cycle the parameter setting through all
the available choices:
PAGE 6.5.2. VPN 98 - 024 FRONT PANEL OPERATION