VOLTECH AT3600 USER MANUAL MEASURE MODE
Because node numbers (1 to 20) are used in the MEASURE Mode, the test nodes are
entered using the keyboard number pad. After each node number has been typed, proceed
to the next by using the cursor (↑ or ↓) soft-keys, or by using the ENTER key ( ↵ ).
If you have have purchased an AT1600 without the optional nodes 11 to 20 fitted, the
display will show node 10 if you try to enter 11 to 20.
For simple tests such as Winding Resistance, there is only one Hi (or Lo) node value to
enter. However, for the ‘C Capacitance’ test, there may be several values for both Hi and
Lo nodes.After you have entered the first Hi node, the display will show:
As you enter more numbers, these will be added to the list of Hi nodes. When you press
the Enter key, or the ↓ soft-key with no new number (for example at the display above),
then the group of Hi nodes will be terminated, and the display will move to the Lo nodes:
When you have entered values for each parameter in the set up list, press the EXIT soft-
key.
FRONT PANEL OPERATION VPN 98 - 024 PAGE 6.5.3.