VOLTECH AT3600 USER MANUAL PROGRAM MODE
Press ‘CONTINUE’ to return to the top level display. Your program group has now
been loaded into the tester’s internal store and the server PC may be disconnected.
These programs will be stored during power off, and will only be overwritten by
another group download.
ERROR MESSAGES
The following is a list of possible error messages, which may occur if there is a fault
in the group download process, together with suggestions for the possible cause.
a)
Possible causes:
Faulty cable: The cable from the tester’s server port to the server PC
has a broken wire, or is not plugged in correctly.
COM port: The COM port has not been set up in the server
application.
Server not running: Either the server PC is not powered on, or the server
application is not running.
b)
Cause:
The directory set up in the server application as the location of test programs
does not contain a group with the requested name.
At all the above displays, pressing the CONTINUE soft key will take you back to the
top level display.
FRONT PANEL OPERATION VPN 98 - 024 PAGE 6.6.5.