SPECIFICATION VOLTECH AT3600 USER MANUAL
PAGE 8.1.2. VPN 98 - 024 SPECIFICATION
Test
User Entered Test Data
Measurement Range Test Signal Test Frequency
(5)
A
R
(1)
WATT
Wattage 1mW to 40W 1V to 270V 20Hz to 1500Hz 0.3%
WATX
Wattage
(External Source)
1mW to 6kW 5V to 600V 20Hz to 5kHz
2
0.3%
SURG
Surge Stress 1mVs to 1kVs 100V to 5kV n/a 3.0%
TRIM
Trimming
Adjustment
n/a n/a n/a n/a
STRW
Stress Wattage 1mW to 40W 1V to 270V 20Hz to 1500Hz 1%
STRX
Stress Wattage
(External Source)
1mW to 6kW 5V to 600V 20Hz to 5kHz
2
1%
R2
DC Resistance
Match
1:1000 to 1000:1 n/a n/a 0.2%
L2
Inductance Match 1:10000 to 10000:1 1mV to 5V 20Hz to 3MHz 0.1%
C2
Inter-winding
Capacitance Match
1:1000 to 1000:1 1mV to 5V 20Hz to 3MHz 0.2%
GBAL
General
Longitudinal
Balance
0dB to 100dB 1mV to 5V 20Hz to 3MHz 0.5dB
LBAL
Longitudinal
Balance
0dB To 100dB 1mV to 5V 20Hz to 3MHz 0.5dB
ILOS
Insertion Loss -100dB to 100dB 1mV to 5V 20Hz to 3MHz 0.5dB
RESP
Frequency Response -100dB to 100dB 1mV to 5V 20Hz to 3MHz 1.0dB
RLOS
Return Loss -100dB to 100dB 1mV to 5V 20Hz to 3MHz 0.2%
Z, ZB
Impedance,
Impedance + bias
1mΩ
to
1M
1mV to 5V 20Hz to 3MHz 0.2%
OUT
Output to User Port n/a n/a n/a n/a
ANGL
Impedance Phase
Angle
-360
°
to
360
°
1mV to 5V 20Hz to 3MHz
0.05
°
PHAS
Inter-winding Phase
Test
-360
°
to
360
°
1mV to 5V 20Hz to 3MHz
0.05
°
ILK
Leakage Current
Test
1µA
to 10mA 1V to 270V 20Hz to 1kHz 0.5%
LSBX
Inductance with
External Bias
(Series)
DC Bias
1nH to 1MH 1mV
1mA
to
to
5V
400mA
20Hz to 3MHz 0.05%
LPBX
Inductance with
External Bias
(Parallel)
DC Bias
1nH to 1MH 1mV
1mA
to
to
5V
400mA
20Hz to 3MHz 0.05%
ZBX
Impedance,
Impedance
+External bias
1mΩ
to
1M
1mV to 5V 20Hz to 3MHz 0.2%
ACRT
Hi-Pot Ranp (AC)
10µA
to 5mA
4
100V to 5.5KV 50Hz to 1kHz 3.0%
DCRT
Hi-Pot Ramp (DC)
1µA
to 3mA 100V to 7KV n/a 3.2%
ACVB
Voltage Break down
(AC)
10µA
to 5mA
4
100V to 5.0KV 50Hz to 1kHz 3.0%
DCVB
Voltage Break down
(DC)
10µA
to 3mA 100V to 7KV n/a 3.2%