VOLTECH AT3600 USER MANUAL SPECIFICATION
SPECIFICATION VPN 98 - 024 PAGE 8.1.3.
Notes:
1. A
R
= The basic relative accuracy (see over for the full specification).
2. Depends on external source type.
3. 650Vrms for AC measurements, or 1000V for DC measurements.
4. Peak value.
5. 1MHz may be the maximum bandwidth for some testers manufactured before 1999. Consult your
supplier for details.
6. Where used, the bias current accuracy is +/-10% of the requested value.
7. Test Frequency Accuracy
a. F <= 16kHz, error is +/-0.25Hz +/- 0.01% of requested frequency
b. F >16kHz and <= 250kHz, error is +/- 4Hz +/- 0.01% of requested frequency
c. F > 250kHz, error is +/- 64Hz +/- 0.01% of requested frequency