Calibration verified by making precise measurements on built-in sample R, L and C
(Verification may be performed at any time).
•
•
•
•
•
•
Verification L and C may be calibrated separately for external traceability.
THE COMPLETE CALIBRATION SYSTEM:
The calibration fixture that fits into the normal position at the top of the tester
(pictured).
PC Software disk to run on a PC (386 or better with 4MB RAM). PC not supplied.
A precision digital multi-meter (not supplied).
An optional IEEE488 interface for the PC for fully automatic calibration
(not supplied).
Fully Automatic Calibration Semi–Automatic Calibration.
Less than 30 minutes About 45 minutes
No manual readings About 30 readings read from the meter and
entered into the PC
IEEE488 Interface required
NI PCII/IIA, IoTech GP488 and CEC suitable
Normal PC Serial ‘COM’ interface required
HP3458A 8½ digit DMM essential HP 3458A DMM or similar required
DMM SPECIFICATION FOR SEMI-AUTOMATIC USE
The Voltech calibration system relies on the accuracy of the DMM used to achieve the
published specification of the tester. The meter must measure:
AC Volts to 0.04% @100Hz and 3% @1MHz
DC Volts to 0.003% @2V and 0.005% @600V
DC resistance using a four terminal connection to 0.05% for 500mΩ and 0.01% for
750Ω.
Keithly 2002 and Datron Wavetek 1271 (with high accuracy AC option) DMMs are
suitable alternative meters.
CALIBRATION ENVIRONMENT
Calibration should be carried out at 23°C ±5°C, relative humidity 10 to 90% non-
condensing.