QUICK START TUTORIAL VOLTECH AT3600 USER MANUAL
PAGE 2.3.6 VPN 98 - 024 GETTING STARTED
(The ‘Ohm’ units button is selected by clicking with the mouse.)
If the "User Offset Enabled" check box is checked, a value can be entered into the
edit box. The value entered (in the units shown) is then added to any results returned
from the AT tester. This function can be used to adjust for measurement fixture
effects that cannot be compensated for or to compensate the fixture manually so a
compensation stage is not required to obtain the correct readings.
Click on the ‘OK’ button. The test and its parameters will now appear in the
‘Program’ window.
Again by double clicking the left mouse button, select ‘R Winding Resistance’ from
the ‘Available Tests’ window.
At the dialogue box, enter the data as before; this time for the second winding:
Integration (Leave as the default - Medium)
High terminal C
Low terminal D
Minimum 59Ω
ΩΩ
Ω
Maximum 73Ω
ΩΩ
Ω
Click on the ‘OK’ button. Again the test and its parameters will appear in the
‘Program’ window.
Now, by double clicking the left mouse button, select ‘LS Inductance (Series
Circuit)’ from the ‘Available Tests’ window.
At the dialogue box, enter the data required for the inductance test:
Signal 1V(again choose the V units button by clicking
with the mouse)
Frequency 50Hz
Integration (Leave as the default - Medium)
High terminal A
Low terminal B
Click on the ‘ > ‘ button to select a minimum limit only, and enter:
Minimum 3 H