VOLTECH AT3600 USER MANUAL QUICK START TUTORIAL
GETTING STARTED VPN 98 - 024 PAGE 2.3.5.
By double clicking the left mouse button, select ‘R Winding Resistance’ from the
‘Available Tests’ window.
The following dialogue box will appear.
Initially enter the terminal names. Input ‘A’ as the high terminal and ‘B’ as the low
terminal, moving between the fill-in boxes using the TAB key.
Now enter the resistance limits. This can be done in four ways:
% Click on this button to enter a nominal value with a percentage tolerance
(for example, 66Ω with 10% tolerance),
>< Click on this button to enter minimum and maximum values (for example,
59Ω and 73Ω),
> Click on this button to enter just a minimum value (for example, > 59Ω),
< Click on this button to enter just a maximum value (for example < 73Ω).
In this example, >< limits will be used: