VOLTECH AT3600 USER MANUAL CREATING THE PROGRAM
TEST PROGRAM EDITOR VPN 98 - 024 PAGE 3.3.19.
4. Have available both a sample transformer, and possibly a shorting header
if you wish to add tests to the program for which you will be applying
short-circuit compensation.
EXAMPLE OF USING THE MEASURE BUTTON WITH THE LS TEST
LS - Winding Inductance (Series Circuit) - is one of the tests where both the test
voltage and frequency, and the actual measurement may be obtained by using the
Measure button. The following example assumes that this is what you wish to do.
Initially select the LS test in the normal way by double-clicking with the mouse in the
Available Tests list.
In the LS test dialogue box:
Enter the terminal names in the normal way.
Select the integration you require if different from the default setting of Medium.
In the Test Parameters section, leave both the Signal and the Frequency edit
boxes blank.
The default units button for the Signal is ‘mV’. With this setting, the Editor will
enter the test voltage chosen by the tester, changing as appropriate to the ‘V’
button if the signal is large. Alternatively, if you wish the test signal to be
determined as a current, click on the ‘mA’ units button.
Finally, in the lower section of the dialogue box, select the desired limits type.
At this point, the dialogue box should be as follows:
Now, using the mouse, click on the Measure button.
Because the LS test is one where short-circuit fixture compensation could be applied,
you will then see the following message: