EasyManua.ls Logo

Voltech AT3600 - Page 92

Default Icon
424 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CREATING THE PROGRAM VOLTECH AT3600 USER MANUAL
PAGE 3.3.20. VPN 98 - 024 TEST PROGRAM EDITOR
If the value of inductance is high, then compensation may not be needed, and you
may click on ‘OK’ with ‘Measure without Fixture Compensation’ selected.
If you click on ‘OK’ with ‘Compensate Fixture’ selected, then there will be an
additional message, which in general depends on whether the test requires short- or
open-circuit compensation. In this case, the LS test requires short-circuit
compensation, so the following dialogue box will be shown:
Fit the required short-circuit header, and click on ‘OK’.
The Editor will now command the tester to perform the Fixture Compensation for the
required test (in this case LS). The tester will then store the ‘zero’ value, and subtract
it from the next result returned for the test.
If the short-circuit links are not correctly fitted, you may see an error message which
allows you to try again, or to cancel the compensation.
Assuming no errors, when the tester has finished the compensation measurement, you
will be prompted to remove the shorting links, and fit the sample transformer:
After you have clicked on the ‘OK’ button, the tester will be commanded to make its
measurements, and you will see the following message: