8 VWR MEGA STAR 3.0 / 3.0R ver b, rel 25.06.2020
Use only with rotors which have been properly installed.
Use only with rotors which have been loaded properly.
Follow the instructions given in the rotor manual.
Never overload the rotor. Follow the instructions given in the rotor manual.
Never open the lid until the rotor has come to a complete stop and this has b
een
c
onfirmed in the display.
T
he lid emergency release may be used in emergencies only to recover the samples
from the centrifuge, e.g. during a power failure.
Never use the centrifuge if parts of its cover panels are damaged or missing.
Do not touch the electronic components of the centrifuge or alter any electronic or
mechanical components.
Please observe the safety instructions.
Please pay particular attention to the following aspects:
Loc
ation: Well-ventilated environment, set-up on a level and rigid surface wit
h
adeq
uate load-bearing capacity.
Rotor installation: Make sure the rotor is locked properly into place before operatin
g
t
he centrifuge.
Especially when working with corrosive samples (salt solutions, acids, bases), the
accessory parts and vessel have to be cleaned carefully.
Always balance the samples.
Centrifuging hazardous substances:
D
o not centrifuge explosive or flammable materials or substances which could react
violently with one another.
The centrifuge is neither inert nor protected against explosion. Never use t
he
c
entrifuge in an explosion-prone environment.
Do not centrifuge inflammable substances.
Remaining risk: Improper use can cause damages, contamination and injuri
es
w
ith fatal consequences.
Do not centrifuge toxic or radioactive materials or any pathogenic micro-organisms
without suitable safety precautions.
When centrifuging microbiological samples from the Risk Group II (according to the
B
io-safety Manualof the World Health Organization (WHO)), aerosol-tight biologica
l
s
eals have to be used.
For materials in a higher risk group, extra safety measures have to be taken.
If toxins or pathogenic substances have gotten into the centrifuge or its parts,
appropriate disinfection measures have to be taken.
Highly corrosive substances which can cause material damage and impair th
e
mec
hanical stability of the rotor, should only be centrifuged in corresponding protectiv
e
t
ubes.