EasyManua.ls Logo

W&H multidem C27 - Installation; Filter Head and Water Connection Assembly

W&H multidem C27
24 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10
Prior to installation, read the technical data (Chapter 9) and the operating and safety information (Chapter 3).
Following storage and transportation below 0 °C, the product must be stored in the open original packaging for at least 24 hours prior to
commissioning at the ambient temperatures for operation specified (Chapter 9).
4. Installation
Assembly the Filter Head, Water Connection
Ensure that the locking handle on the head is open.
Close water intake and switch off the power supply to the terminal device.
Fit the hoses to the water inlet IN and outlet OUT on the filter head .
Do not install the intake hose if there is no non-return valve on the water inlet IN.
Ensure that the flat seals supplied are used.
The max. tightening torque at the G3/8” connections must not exceed 14 Nm! Only hose connections with flat
seals may be used, as hoses with conical connections damage the filter head connections and invalidate any
guarantee claims!