Questions and Answers

  • A
    anne76Aug 17, 2025
    How to fix a W&S Lexa MN-111 Laboratory Equipment sterilizer that remains switched OFF?
    • C
      christianleblancAug 18, 2025
      If your W&S Laboratory Equipment sterilizer remains switched off, there could be several reasons. First, check if the main switch or network circuit breaker is turned off and activate it. Second, verify that there is voltage at the socket. Finally, ensure that the power cord is correctly connected.
  • R
    Rhonda RodriguezAug 20, 2025
    Why are there corrosion or spots on instruments in my W&S Laboratory Equipment?
    • M
      Melanie MitchellAug 20, 2025
      Corrosion or spots on instruments in your W&S Laboratory Equipment can occur due to several reasons. Ensure instruments are dry before placing them in the sterilizer. Use water of good quality and drain both water tanks. Clean, rinse, and dry instruments to remove organic or chemical residues. Clean the chamber, trays, and chamber rack. Avoid contact between instruments of different materials. Also, clean the chamber to remove scale deposits and use good quality water.
  • M
    Matthew CaseAug 22, 2025
    What to do if there is residual water in the chamber of my W&S Laboratory Equipment after the cycle?
    • Y
      yfryAug 22, 2025
      If you find residual water in the chamber of your W&S Laboratory Equipment after a cycle, it could be due to several factors. Make sure the sterilizer is properly leveled. Avoid overloading the chamber and comply with the maximum load weight limits, always using the chamber rack for trays and cassettes. Check if the chamber filter is clogged and clean it if necessary. Ensure all chamber filter caps are well-positioned. Also, make sure the load is placed correctly.
  • K
    Keith MitchellAug 25, 2025
    Why is water leaking at the front of my W&S Laboratory Equipment sterilizer?
    • T
      Timothy JacksonAug 26, 2025
      If water is leaking at the front of your W&S Laboratory Equipment sterilizer, it could be due to leaks through the chamber door gasket. In this case, clean or replace the door gasket and clean the chamber face side. If the leak is internal, call technical service.
  • R
    rkleinAug 27, 2025
    Why does the chamber door of my W&S Laboratory Equipment lock and then immediately re-open?
    • P
      Phillip WebsterAug 27, 2025
      If the chamber door of your W&S Laboratory Equipment locks and then immediately re-opens with an “Open the door” message, it might be due to a door gasket not properly placed or a seal sticking out, so ensure that the door gasket is evenly inserted on the entire circumference. Alternatively, the door might be jammed by external objects or the load itself. Remove any objects interfering with the chamber door and ensure the door does not force against the load or the chamber furniture.
  • D
    Dustin MartinAug 30, 2025
    What to do if the cycle starts on my W&S Lexa MN-111 Laboratory Equipment but there is no pressure or temperature rise?
    • N
      Natalie DanielsAug 30, 2025
      If the cycle commences on your W&S Laboratory Equipment but there is no pressure or temperature rise, the safety thermostat switch might be open. Reset the safety thermostat switch. If the issue persists, it could be an electric-electronic fault, in which case call technical service.
  • M
    Marisa TorresSep 1, 2025
    Why isn't the automatic water filling working in my W&S Lexa MN-111?
    • S
      Steven DennisSep 1, 2025
      If your W&S Laboratory Equipment is connected to an automated water supply system and the automatic water filling isn't working, first ensure the water fill system is connected to the sterilizer. If the water fill system attempted to fill the tank when water was temporarily unavailable, switch the sterilizer off and then on again, check the external water supply system, and check for water leaks from the sterilizer. A faulty MIN water level sensor in the clean water tank could also be the reason, in which case call service.
  • S
    Samuel AtkinsonSep 4, 2025
    Why are instruments turning brown or black in my W&S Lexa MN-111?
    • J
      John RasmussenSep 4, 2025
      If instruments are turning brown or black in your W&S Laboratory Equipment, it could be due to an incorrect temperature selected. Select a sterilization cycle featuring a lower sterilization temperature and follow the instructions of the instrument manufacturer.
  • W
    Wayne MartinezSep 7, 2025
    Why is the cycle report printer not working with my W&S Lexa MN-111 Laboratory Equipment?
    • A
      Anthony BakerSep 7, 2025
      If the cycle report printer is not working with your W&S Laboratory Equipment, check the data and the power connection to the printer.
  • A
    Alex WeaverSep 9, 2025
    What does the USB saving warning mean on my W&S Lexa MN-111?
    • K
      Kenneth JohnsonSep 9, 2025
      If you see a warning about USB saving (HTML and SCL files) on your W&S Laboratory Equipment, it means the USB pen drive is not connected or not properly connected to the sterilizer. Check the presence and connection of the USB pen drive.

Summary

Instructions for Use

Safety Information

Safety Warnings and Risks

Covers thermal, electrical, and misuse hazards associated with the sterilizer.

Getting Started

Installation Procedures

Guidance on location, electrical, and water connections for proper installation.

Administrator Functions

User Management

Procedures for managing user accounts, including adding, deleting, and resetting PINs.

Sterilization Cycles

Load Preparation and Placement

Instructions for cleaning instruments, correct load placement, and types of loads.

Sterilization Cycle Management

Covers running cycles, setting start times, drying times, and cycle completion.

Maintenance

Maintenance Warnings and User Tasks

Safety precautions and user-level maintenance tasks like cleaning.

Diagnostics

Error Codes and Troubleshooting

Lists error codes, their causes, actions, and general troubleshooting advice.

Emergency Door Opening

Detailed steps for safely opening the sterilizer door in emergency situations.

Technical Data

Sterilization Programs and Phases

Information on standard/custom programs, their parameters, and cycle phases.

W&S Lexa MN-111 Specifications

General IconGeneral
CategoryLaboratory Equipment
ModelLexa MN-111
ManufacturerW&S
Power Supply220V / 50Hz
MaterialStainless Steel