EasyManua.ls Logo

Weidmüller UHZ/AR - Output Calibration Procedure

Default Icon
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
22
Equipment requirements
• An accurate digital multimeter (accurate to 0.05 mV and ±0.1 μA)
Output calibration procedure
When the
display shows
Action/Description
ì Å
ENT
Allow fifteen minutes of powered operation (to reach a stable temperature).
Put the instrument in setup mode and scroll through the main menu
COPn
Press *.
COPy
Calibrate Outputs? Toggle Start
Measure the output current
I=0
Press *. The output current should be 0.00 mA
(Use the keypad to adjust the value as required)
Increase Decrease Accept
I=20
Press *. The output current should be 20.00 mA
Increase Decrease Accept
Measure the output voltage
E=0
Press *. The output voltage should be 0.00 V
Increase Decrease Accept
E=10
Press *. The output voltage should be 10.00 V
Increase Decrease Accept
Save
Instrument will save values and return to normal operation (using the new values).
Note: Do not remove the power while the save message is on display.