22 Chapter 3 Operation Welch Allyn 9600 Plus Calibration Tester
12.Wait one full minute before taking another reading with the
same thermometer. Repeated measurements in short
sequence may cause higher readings.
13.When using only one 9600 Plus Calibration Tester, test all
available thermometers for calibration verification at the
current calibration set point temperature before proceeding to
the next calibration set point temperature. See "Changing the
Calibration Set Point Temperature" (page 11).
14.The device will exit CAL mode after 10 minutes of inactivity, or
can be forced to exit by holding both C/F and Mem for 5
seconds. After exiting Calibration Check Mode the
thermometer will return to Sleep Mode.
The probe cover must be replaced with a new probe
cover before each reading, to remove any residual heat
absorbed from the 9600 Plus Calibration Tester.
Do not hold the thermometer while waiting. This allows
the thermometer to remain at ambient room
temperature.