Directions for Use Chapter 4 Maintenance 31
Specifications
Power
Requirements
Voltage: 100 - 240 V 50/60 Hz
Current Heating: 480 mA
After Warm-Up: 185 mA
Power
Consumption
Heating: 2.4 W
After Warm-Up: 0.92 W
Calibration Tester Length: 254 mm (10.00”)
Width: 105 mm (4.13”)
Height: 85 mm (3.34”)
Weight (without Power Pack): 0.75 kg (1.65 lbs)
Power Pack
(approximate
dimensions)
Length: 86 mm (3.38”)
Width: 51 mm (2.00”)
Height (with plug inserted): 62 mm (2.44”)
Weight: 0.18 kg (0.40 lbs)
Environment Storage Temperature: 0°C – 50°C (32°F – 122°F)
Ambient Operating Temperature: 18.3° C to 26.7°C (65° F to 80° F)
Operating Humidity: 20% to 95% RH. Non-Condensing
Operating
Specifications
Controlled Temperature of 9600 Plus
Low Setting: 36
o
C (96.8
o
F)
Intermediate Setting: 38.5
o
C (101.3
o
F)
High Setting: 41
o
C (105.8
o
F)
Temperature Accuracy: ±0.05° C (0.1
o
F)
Maximum Pressure/Altitude: <
2000 meters.
Heating and
Cooling Times
Tested at an ambient
of 23.9°C (75°F)
Room Temperature to Low Setting: Approximately 20 min.
Low Setting to Intermediate Setting: Approximately 6 min.
Intermediate Setting to High Setting: Approximately 6 min.
Cool from High Setting to Low Setting: Approximately 25 min.
Installation Category II