Guidance and manufacturer’s declaration:
Electromagnetic immunity (IEC 60601-1-2 Table 202)
The Welch Allyn AED 10 Trainer is intended for use in the electromagnetic environment
specified below. The customer or the user of the Welch Allyn AED 10 Trainer should
assure that it is used in such an environment.
Immunity Test IEC 60601 test level Compliance
level
Electromagnetic environment
- Guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast transient/
burst IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input/output lines
Not applicable
Not applicable
Surge
IEC 61000-4-5
± 1 kV differential mode
+/- 2 kV common mode
Not applicable
Not applicable
Voltage dips, short
interruptions and voltage
variations on power supply
input lines.
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
) for 5 cycles
70% U
T
(30% dip in U
T
) for 25 cycles
<5% U
T
(>95% dip in U
T
) for 5 sec
(U
T
is the a.c. mains voltage prior to
application of the test level.)
Not applicable
Not applicable
Not applicable
Not applicable
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m Not applicable Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment