LAMP REPLACEMENT
To remove old lamp:
1. Turn AudioScope 3 “OFF.”
2. LAMP MAY BE HOT—ALLOW TO COOL
BEFORE PROCEEDING.
3. Follow instructions for cover removal (page 14).
4. Depress the blue lamp ejector lever (see Figure 1 below). The
lamp will pop up.
5. Grasp lamp and remove.
Fiber Optic Bundle
Beveled end of lamp
Lamp ejector lever
Figure 1
To insert new lamp:
1. For best performance of Halogen lamps, remove any dirt or
fingerprints from glass with a clean cloth moistened with 70%
Isopropyl Alcohol before inserting lamp. Failure to follow clean-
ing procedures will result in permanent stains on the glass with
reduction in light output. Average lamp life is 20 hours of on-
time.
2. Rest lamp on lamp holder bracket with beveled end of lamp
toward fiber optic bundle (see Figure 1 above).
3. Push lamp into lamp holder.
4. Follow instructions for cover replacement (page 16).
NOTE: Replace with Welch Allyn No. 06200 lamp only. Other
lamps may cause damage to the instrument or may cause
inaccurate readings.
15