This setting (PRINT-AUD) selects the format for the
printout only. An audiogram always appears on the
screen while in this mode.
NORMAL BOX FORMAT
It is possible to have the normal box, as described by
ASHA,
appear on the
tympanogram screen and printout. The boundaries for this normal box are -150
daPa
to
+100
daPa
and 0.2
cm
3
to 1.4
cm
3
.
NOTE
A compliance value of 7.5
cm
3
or greater will
automatically turn off the
ASHA
normal box.
The normal box is the default setting. To deselect this normal box, move the square
cursor with either the
<
or
>
Hz so that it is placed in front of the words Normal Box Off.
While the cursor is in this position, select the M+ pushbutton to save this feature as the
new default setting. Note that the word SAVED appears in the lower right-hand margin.
This message assures you that the normal box ill not appear on the tymp screen or
printout.
AUDIOGRAM RANGE
All eleven frequencies are available during audiometry or the range can be abbreviated
to eight frequencies.
The default setting is Aud Range Normal. To select the
abbreviated frequency range, position the square cursor in front of the feature Aud
Range Narrow. Press the M+ pushbutton to save this narrow range for audiometric
testing. Note that the word SAVED will appear in the lower right-hand margin and the
asterisk now appears in front of the narrow range selection. The normal range of
frequencies include 125 Hz through 8000 Hz. The narrow range of frequencies
includes 500 Hz through 6000 Hz. Please note that in the Aud mode, if the narrow
range is selected, the
<
and
>
Hz pushbuttons will allow you to scroll through this
abbreviated frequency range only. Both the screen and printout will still be
labelled
with the full range of frequencies. i.e., 125 Hz through 8000 Hz.
EXIT PROGRAM MODE
Exit the program mode by selecting the PROG pushbutton. Note that you return to the
test mode which was operational prior to entering the program mode.