is performed by the 226 Hz band pass filter (U65). The filtered RMS microphone signal
is
rectified
to a dc level for input to the Microprocessor A/D convertor.
This filtered signal
is
level adjusted (U59) by the Microprocessor to establish an optimum operating range for
the system during the tymp sweep or reflex testing.
C
i
i
34