CAUTION: DISCONNECT LS200 FROM POWER SOURCE BEFORE SERVICING.
I YOKE R & R, CONTINUED.....
I 18. TEST JOINTS.
WITH THE LS200 IN AN UPRIGHT POSITION, CYCLE THE ARM UP AND
DOWN THREE TO FIVE TIMES. CHECK FORT AN;Y BINDING OF THE
JOINT. WITH THE ARM AT 45 DEGREES, CHECK FOR DROOP. WITH THE
ARM ALL THE WAY DOWN AT THE 90 DEGREE POSITION, CHECK FOR
SPRINGBACK. TEST TO VERIFY SAFE OPERATION. CONNECT TO A.C.
POWER SOURCE AND VERIFY THAT ALL THREE LAMPS ILLUMINATE.