APPENDIX EE
PRODUCT DESCRIPTION LS200
PROCEDURE LIGHT
The LS200 is a halogen procedure light with two mounting options...mobile stand and wall
mount. It is used for routine medical examinations and as such can be positioned in a variety of
positions. The light housing is manufactured of injection molded plastic which minimizes surface
temperature. Light is provided by three quartz halogen lamps rated at 12 volts and 20 Watts each.
The power source for the LS200 is standard A.C. for the country it is sold into. Two power sources
are available for the LS200: Domestic 120 VAC/60 Hz and International 240 VAC/50Hz. The
secondary, or lamp voltage is 12 VAC. All models use the same 20 watt halogen lamp PN 06400.
Lamp current is switched by a Double Pole Single Throw (DPST) switch. The transformer is
energized whenever the power cord is plugged into an active AC outlet (when fuse drawer contains
two good fuses).
Please refer to the Owner’s Manual PN 441118-1ML “Set-Up and operation” to become
familiar with the LS200.
This product is designed and manufactured to strict specifications and safety regulations for
medical products, and should be serviced by Welch Allyn trained technicians only.
44

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Welch Allyn LS200 Procedure Light and is the answer not in the manual?

Welch Allyn LS200 Procedure Light Specifications

General IconGeneral
BrandWelch Allyn
ModelLS200 Procedure Light
CategoryLighting Equipment
LanguageEnglish

Related product manuals