Apply sample to prism, close plate, and let adjust to ambient temperature for 30 seconds.
Hold to light, look into eyepiece, focus to see graduations; upper field blue, lower white.
Turn Calibration Screw until boundary meets 0.0° mark for ambient temperature calibration.
Apply 2-3 drops of liquid sample to prism, close plate, and let adjust to ambient temperature.
Hold to light, look into eyepiece, focus to see graduations; upper field blue, lower white.
Read the boundary line where blue and white fields meet the graduated scale for Brix value.
Dilute lubricant using accurate measuring cup to create a 'Gold Standard' solution for target reading.
Measure the 'Gold Standard' solution with the refractometer to determine the target reading.
Record the measured reading of the 'Gold Standard' for future reference.
Use refractometer to ensure new dilutions match the 'Gold Standard' reading for accuracy.
Accurate measurement requires calibration and sample adjustment to ambient room temperature.
Avoid damp conditions and do not immerse instrument in water to prevent fogging.
Clean prism with a soft, damp cloth regularly to prevent inaccurate results and damage.
Do not measure abrasive or corrosive chemicals as they can damage the prism's coating.
Handle and store this optical instrument carefully to prevent damage to components and structure.