WIKA operating instructions models TG53, TG54
When mounting a bimetal thermometer that can be rotated and
swivelled, the specic instructions must be followed. The following steps
must be taken in order to set the indicator to the desired position:
1. Loosen the lock nut or union nut at the process connection.
2. Loosen the hexagon bolts and slotted screws at the swivel joint.
3. Position the indicator as required, tighten the hexagon bolts and
slotted screws, and nally tighten the lock nut or union nut rmly.
loosening
Make sure to loosen the
screws on the opposite side
as well!
Installation with
open-ended spanner
5. Commissioning, operation