Fault Possible cause Solutions
The scaling tip doesn't
vibrate but there is water
owing out when stepping
on the switch.
The tip is in loose contact. Screw the tip on the
handpiece tightly (See Figure
6).
The connect plug between the
handpiece and the circuit board
is in loose contact.
Contact our dealers or us.
Something wrong with the
handpiece.
Send the handpiece to our
company to repair.
Something wrong with the
cable.
Contact our dealers or us.
The scaling tip vibrates
but there is no spray
when stepping on the foot
switch.
The water control knob is not
on.
Turn on the water contrl
knob [note 1].
The vibration of the tip
becomes weak.
The tip hasn't been screwed on
to the handpiece tightly.
Screw the tip on the
handpiece tightly (See Figure
6).
The tip is loose because of
vibration.
Screw on the tip tightly (See
Figuree 6).
The coupling between the
handpiece and the cable isn't
dry.
Dry it by the hot air.
The tip is damaged [note 2]. Change a new one.
There is water seeping
from the coupling between
the handpiece and cable.
The waterproof "O"ring is
damaged.
Change a new waterproof
"O"ring.
The u-le doesn't vibrate. The screw is loose. Tighten it.
Endochuck is damaged. Change a new one.
There is noise coming
from the endochuck.
The screw is loose. Tighten it.