EasyManua.ls Logo

WTC Technitron T2050 - Primary Current Calibration Procedure

WTC Technitron T2050
61 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
BUSQ\YRbQdY^WdXUD" %
DUSX^Ydb_^D" % GU \T3_^db_\
FUbcY_^F! !Q^TF! !1
#'
'! "$$
3QedY_^*
>5F5BQddU]`dd_SQ\YRbQdU`bY]Qbi_bcUS_^TQbi
SebbU^dY^3_^cdQ^d3ebbU^d]_TU1\gQicSXUS[V_b
dXU`b_`UbSebbU^dbQ^WUVYbcdRiTU\YfUbY^WQd))
Y^1F3]_TUQ^T]UQcebY^WdXUSebbU^d>5F5B
cgYdSXd_3_^cdQ^d3ebbU^d]_TUgYdX_edVYbcd
SXUS[Y^WdXUS_^db_\µcQRY\Ydid_QSSebQdU\ibU`_bd
SebbU^dQd" % Q^T))1F3SebbU^d
Remember: If any range jumper is changed the control is NOT
calibrated in the new range. If the jumper is returned to the
original calibrated position, the control is still NOT calibrated.
Calibrated current values reside in two memory locations, one for
primary and one for secondary. The control does not "know" or
remember where the range jumpers are located.
3ULPDU\&XUUHQW&DOLEUDWLRQ3URFHGXUH
To calibrate primary current, follow this procedure:
1. Choose a primary current range (W1 W4) that cannot be
exceeded at a 99% available current weld.
2. Turn off power and set SW11 to be OPEN (AVC Mode).
3. Program a simple weld schedule:
SQUEEZE 20 CY
WELD TIME 15 CY (TURNS RATIO = 1)
WELD CUR 99%
HOLD 10 CY
OFF 0 CY
4. Set all other functions off (set to 0), EXCEPT Impulses and
Turns Ratio = 1.
5. Make a weld, with or without work. Measure the secondary
current with a secondary current meter such as Technitrons
WS10, blanking the first 3 cycles in the measurement on the
WS10. (First = 4, last = 15 on the WS10 or WS25.)

Related product manuals