2.1.2. Mounting of the sensors / codes
The following points must be observed:
2.1.2.1. Correct horizontal position
Sensor-Type
Codes:
XA-XXX Sensor type 1 Arcdeg.
XB-XXX Sensor type 5 Arcdeg.
XC-XXX Sensor type 10 Arcdeg.
XD-XXX Sensor type 30 Arcdeg.
XE-XXX Sensor type 60 Arcdeg.
The sensors are to be mounted so that the reading at the flange horizontally correct. The meaning of the coding is
the following.
The sensor measuring range is marked with the letters A to E. The other numbers are part of the identification
code
2.1.2.2. Angular position
The positive or negative inclination is according to the following definition (Looking at the sensor with the cable
connector to the right):
Inclination: positive
Inclination: horizontal (0 Degree)
Inclination: negative
ATTENTION / IMPORTANT for measurements with ZEROTRONIC sensors
For precise measurements it is important that the sensors have been in operation for at least 15 minutes (start
up time) before the first measurement with a LEVELMETER 2000.
After putting it in function the sensor is slightly heated up by some of the built-in electronic components. After the
start-up time mentioned above the system is stabilised and ready for measurements.
Disabling the automatic shut-off of the LEVELMETER 2000
When pressing the
ON/MODE key for more than 3 seconds the display starts flashing and the automatic shut
off is disabled. In the standard mode the instrument shut off after about 10 minutes.
Exeption: In case the LEVELMETER 2000 is powered by a connection to a mains adapter, the instrument is
never shutting off automatically.