OXYGEN CONCENTRATOR USER MANUAL
14
TECHNICAL PARAMETERS
Model CP101
Working Voltage
230 V
~
± 10% / 50 Hz ± 1 Hz
Rated Power (VA)
770
Output Pressure=0/7kPa,
the flow range(L/min)
1~10
Output Pressure=0,the
purity value(within 10min,
up to the standard)
Oxygen flow=1~10 L/min,purity≥90%
Highest Flow 10 L/min
Highest Flow with 7kPa
pressure
,
change of flow
≤
1 L/min
Highest Flow, purity
(
within
10min
,
up to the standard
)
≥
90%
Range of adusting flow
0~10L/min adusted continously
Net Weight(Kg)
20.5
Noise dB(A)
≤55
Dimension
(
mm
)
L400 × W300 × H540
Oxygen output pressure 60kPa-80kPa
Yellow low oxygen indicator
light
When 72%≤purity≤82%,the yellow indicator will
shin,contact the supplier, Make sure have the spare
oxygen nearby when use it continously.
Red alarm indicator light
When purity ≤72%,red indicator will alarm,and
dispaly show Lo-o2.It will stop within one minute.Please
turn off at once and choose the spare oxygen , contact
the supplier